Purpose: To evaluate the early clinical results achieved with an algorithm adjusted for an expanded range of correction in wavefront-guided customized ablation with the LADARVision4000 (Alcon Laboratories Inc, Fort Worth, Tex).

Methods: Fifty-five consecutive eyes from 31 patients underwent wavefront-guided, customablation laser in situ keratomileusis (LASIK) (Hansatome and BD 4000 microkeratomes, LADARWave aberrometer, LADARVision4000 laser system). These were normal myopic and astigmatic eyes that had never been operated on. The spherical equivalent refractive error was +0.30 to -8.13 diopters (D) (mean -4.26 +/- 2.14 D) and the astigmatism ranged from 0 to -3.75 D (mean -0.97 +/- 0.96 D). eyes were treated according to the CustomCornea protocol using the commercial LADARVision4000 platform. Patients were followed for at least 2 months and standard visual measurements were taken and recorded at 1 day and 1, 3, and 6 months after treatment. The results of treatment were assessed using the following parameters: uncorrected visual acuity (UCVA) and best spectacle-corrected visual acuity (BSCVA); manifest spherical equivalent refractive error; wavefront measurement of high order aberrations; and subjective reports of visual symptoms by patients.

Results: One month after LASIK, the mean manifest spherical equivalent refractive error was -0.18 +/- 0.41 D. Seventy-eight percent and 96% of eyes had manifest spherical equivalent refractive error within +/-0.50 D and +/-1.00 D of attempted correction, respectively, with 78% of eyes with 20/20 or better UCVA. Eyes with 20/16 and 20/12.5 BSCVA were 89% and 36%, respectively (55% and 7% preoperatively). The mean high order aberrations root-mean-square was 0.34 +/- 0.10 microm preoperatively and 0.35 +/- 0.09 microm postoperatively (6.0-mm pupil size). High order aberrations were either reduced, unchanged, or increased by <10% in 58% of eyes. Subjectively, patients reported no visual symptoms and had no complaints regarding the quality of their vision.

Conclusion: The new optimized algorithm for higher refractive errors appears to be effective in improving BSCVA, and a minority of patients demonstrated an increase in the magnitude of high order aberrations when compared to preoperative aberrometry. A slight spherical equivalent refraction undercorrection will be addressed with future nomogram adjustments.

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081-597X-20040901-40DOI Listing

Publication Analysis

Top Keywords

spherical equivalent
16
equivalent refractive
16
refractive error
16
manifest spherical
12
high order
12
order aberrations
12
expanded range
8
visual acuity
8
eyes
6
+/-
5

Similar Publications

Background: The aim of this study was to evaluate visual outcomes and patient satisfaction after bilateral implantation of a new hydrophobic acrylic intraocular lens called Clareon (Alcon) using the mini-monovision technique.

Methods: A single-center, prospective, nonrandomized study was conducted in Tandil (Buenos Aires, Argentina), including patients scheduled for cataract surgery. To achieve mini-monovision, the spherical equivalent was calculated between -0.

View Article and Find Full Text PDF

The rising prevalence of myopia is a significant global health concern. Atropine eye drops are commonly used to slow myopia progression in children, but their long-term use raises concern about intraocular pressure (IOP). This study uses SHapley Additive exPlanations (SHAP) to improve the interpretability of machine learning (ML) model predicting end IOP, offering clinicians explainable insights for personalized patient management.

View Article and Find Full Text PDF

The lipid content of nine dinoflagellates was analyzed using flow cytometry to compare lipid levels. Additionally, the correlation between lipid content, cell size, and carbon content in dinoflagellates was evaluated using BODIPY 505/515 staining. The flow cytometry side scatter (SSC) effectively represented relative cell size, showing a linear relationship with the equivalent spherical diameter (ESD).

View Article and Find Full Text PDF

Spherical implantable collamer lens with postoperative adjunctive LASIK in the treatment of high compound hyperopic astigmatism.

BMC Ophthalmol

January 2025

Department of Ophthalmology and Visual Sciences, McGill University, 5252 Maisonneuve Blvd W, 4th floor,, Montreal, Québec, H4A 3S5, Canada.

Objective: To assess refractive and visual outcomes of a spherical Implantable Collamer Lens (ICL) followed by planned postoperative adjunctive laser-assisted in situ keratomileusis (LASIK) in the treatment of high compound hyperopic astigmatism.

Methods: In this prospective, multi-center, multi-surgeon, single-arm trial, eyes with ≥ 3.50 D hyperopia and ≥ 2.

View Article and Find Full Text PDF

The Influence of Axial Length on Visual Quality following SMILE Surgery in Patients with High Myopia.

Photodiagnosis Photodyn Ther

January 2025

Department of Ophthalmology, Shanghai Tenth People's Hospital, School of Medicine, Tongji University Shanghai, 200072, China.. Electronic address:

Purpose: This retrospective study aimed to investigate the subjective and objective visual outcomes following Small Incision Lenticule Extraction (SMILE) surgery in high myopic patients with varying axial lengths (AL).

Methods: The study enrolled 113 highly myopic patients (202 eyes) who underwent SMILE surgery at Shanghai's Tenth People's Hospital from July 2021 to September 2023. Patients were classified into three groups based on the axial length before surgery: Group A (AL < 26mm, 62 eyes), Group B (26mm ≤ AL < 27mm, 88 eyes), and Group C (AL ≥ 27mm, 52 eyes).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!