Background: Accurate alignment of the components in total knee arthroplasty is important. By use of postoperative CT controls, we studied the ability of a robotic effector to accurately place and align total knee arthroplasty (TKA) components according to a purely CT-based preoperative plan.
Patients And Methods: Robotic TKA was performed in 13 patients (6 men) with primary gonarthrosis. Locator screws were placed into femur and tibia under spinal anesthesia. A CT-scan including the femoral head, knee and ankle was performed. In the preoperative planning software, virtual components were positioned into the CT volume. In a second operation, the robot milled femur and tibia with a high-speed milling tool according to the preoperative plan. On the 10th day, CT controls were performed following the same protocol as preoperatively.
Results: The mean deviation of the postoperative from the preoperatively planned mechanical axis was 0.2 degrees (95% CI: -0.1 degrees to 0.5 degrees ). The accuracy of angular component placement in frontal, sagittal and transverse planes was within +/-1.2 degrees , and the accuracy of linear component placement in mediolateral, dorsoventral and caudocranial directions was within +/-1.1 mm.
Interpretation: Robotic TKA allows placement of components with unparalleled accuracy, but further development is mandatory to integrate soft-tissue balancing into the procedure and make it faster, easier and cheaper.