A new mechanical testing device for measuring anteroposterior knee laxity.

Am J Sports Med

Department of Orthopaedic Surgery, Hôpital Cantonal, Fribourg, Switzerland.

Published: January 2005

Background: The authors assessed a new instrument, the Rolimeter, for the measurement of anteroposterior translation in the knee; it was compared to the KT-1000 arthrometer.

Purpose: To determine if the Rolimeter offers a valid method for the measurement of anteroposterior translation that is as reproducible and reliable as the KT-1000 arthrometer.

Methods: Two of 3 observers examined 16 normal subjects (32 knees) and 36 patients (72 knees) with ligament ruptures twice, using both the Rolimeter and the KT-1000 arthrometers, 30 minutes apart. Total anteroposterior translation (manual maximal Lachman test) was recorded at 20 degrees and 80 degrees of knee flexion.

Results: On average, the Rolimeter measured approximately 1 mm less anteroposterior displacement than the KT-1000 arthrometer at manual maximum stress. Rolimeter measurements were more consistent than the KT-1000 measurements as measured by our observers. Specificity and sensitivity were equivalent between the Rolimeter and the KT-1000 arthrometer.

Conclusions: The Rolimeter is as reproducible and reliable as the KT-1000 arthrometer. It offers a valid method for the measurement of anteroposterior translation in the knee. Higher accuracy was obtained at 20 degrees of flexion for the KT-1000 arthrometer and at 80 degrees of flexion for the Rolimeter.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0363546504267050DOI Listing

Publication Analysis

Top Keywords

anteroposterior translation
16
measurement anteroposterior
12
kt-1000 arthrometer
12
rolimeter
8
translation knee
8
kt-1000
8
offers valid
8
valid method
8
method measurement
8
reproducible reliable
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!