Our purpose in this study was to evaluate the fundamental image quality characteristics of a new slot-scan digital chest radiography system (ThoraScan, Delft Imaging Systems/Nucletron, Veenendaal, The Netherlands). The linearity of the system was measured over a wide exposure range at 90, 117, and 140 kVp with added Al filtration. System uniformity and reproducibility were established with an analysis of images from repeated exposures. The modulation transfer function (MTF) was evaluated using an established edge method. The noise power spectrum (NPS) and the detective quantum efficiency (DQE) of the system were evaluated at the three kilo-voltages over a range of exposures. Scatter fraction (SF) measurements were made using a posterior beam stop method and a geometrical chest phantom. The system demonstrated excellent linearity, but some structured nonuniformities. The 0.1 MTF values occurred between 3.3-3.5 mm(-1). The DQE(0.15) and DQE(2.5) were 0.21 and 0.07 at 90 kVp, 0.18 and 0.05 at 117 kVp, and 0.16 and 0.03 at 140 kVp, respectively. The system exhibited remarkably lower SFs compared to conventional full-field systems with anti-scatter grid, measuring 0.13 in the lungs and 0.43 in the mediastinum. The findings indicated that the slot-scan design provides marked scatter reduction leading to high effective DQE (DQEeff) of the system and reduced patient dose required to achieve high image quality.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.1783531 | DOI Listing |
J Med Imaging (Bellingham)
July 2016
University at Albany , State University of New York, Department of Physics, 1400 Washington Avenue, Albany, New York 12222, United States.
Conventional mammography can suffer from poor contrast between healthy and cancerous tissues due to the small difference in attenuation properties. Coherent scatter slot scan imaging is an imaging technique which provides additional information and is compatible with conventional mammography. A Monte Carlo simulation of coherent scatter slot scan imaging was performed to assess its performance and provide system optimization.
View Article and Find Full Text PDFNihon Hoshasen Gijutsu Gakkai Zasshi
March 2012
Department of Radiology, Kanazawa University Hospital.
The purpose of this study was to optimize X-ray conditions for full spine X-ray examinations in slot-scan digital radiography (SSDR). Follow-up of spinal deformities, such as scoliosis, typically involves many radiographs of the patient throughout childhood and adolescence. The Radiation doses for the full spine X-ray examination should be minimized.
View Article and Find Full Text PDFMed Phys
September 2004
Digital Imaging Research Division, Department of Radiology, Duke University, DUMC 3302, Durham, North Carolina 27710, USA.
Our purpose in this study was to evaluate the fundamental image quality characteristics of a new slot-scan digital chest radiography system (ThoraScan, Delft Imaging Systems/Nucletron, Veenendaal, The Netherlands). The linearity of the system was measured over a wide exposure range at 90, 117, and 140 kVp with added Al filtration. System uniformity and reproducibility were established with an analysis of images from repeated exposures.
View Article and Find Full Text PDFEur Radiol
September 2004
Department of Radiology, University Hospital Charité, Schumannstrasse 20/21, 10117 Berlin, Germany.
X-ray spectra are composed of a broad bremsspectrum and anode-characteristic emission lines. In mammography typically molybdenum (Mo), rhodium (Rh) or tungsten (W) anodes are used in combination with Mo, Rh or aluminium filters. Only the photons with energies between 17 and 22 keV of the resulting spectrum are suitable for the soft tissue imaging needed for mammography.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!