Determination of tiapride in human plasma using hydrophilic interaction liquid chromatography-tandem mass spectrometry.

Arch Pharm Res

Drug Metabolism and Bioanalysis Laboratory, College of Pharmacy and Phytofermentation Research Center, Wonkwang University, Iksan 570-749, Korea.

Published: September 2004

A rapid, sensitive and selective hydrophilic interaction liquid chromatography-tandem mass spectrometric (HILIC-MS/MS) method for the determination of tiapride in human plasma was developed. Tiapride and internal standard, metoclopramide were extracted from human plasma with dichloromethane at basic pH and analyzed on an Atlantis HILIC silica column with the mobile phase of acetonitrile-ammonium formate (190 mM, pH 3.0) (94:6, v/v). The analytes were detected using an electrospray ionization tandem mass spectrometry in the multiple-reaction-monitoring mode. The standard curve was linear (r=0.999) over the concentration range of 1.00-200 ng/mL. The coefficient of variation and relative error for intra- and interassay at three QC levels were 6.4-8.8% and -2.0-3.6%, respectively. The recoveries of tiapride ranged from 96.3 to 97.4%, with that of metoclopramide (internal standard) being 94.2%. The lower limit of quantification for tiapride was 1.00 ng/mL using 100 microL of plasma sample.

Download full-text PDF

Source
http://dx.doi.org/10.1007/BF02975840DOI Listing

Publication Analysis

Top Keywords

human plasma
12
determination tiapride
8
tiapride human
8
hydrophilic interaction
8
interaction liquid
8
liquid chromatography-tandem
8
chromatography-tandem mass
8
mass spectrometry
8
internal standard
8
plasma
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!