In many medical procedures where accurate radiation dose measurements are needed, the variation of detector response with x-ray energy is of concern. The response of LiF:Mg,Cu,P TLDs to a range of x-ray energies was analysed in monoenergetic (synchrotron), diagnostic and therapy radiation beams with the aim of implementing this dosimeter into clinical practice where existing dosimetry techniques are limited due to lack of sensitivity or tissue equivalence (e.g. neonatal radiography, mammography and brachytherapy). LiF:Mg,Cu,P TLDs in different forms from two manufacturers (MCP-N: TLD Poland, GR200: SDDML China) were irradiated using x-ray beams covering 10 keV to 18 MVp. Dose readings were compared with an ionization chamber. The effect of different TLD types and annealing cycles on clinical utility was investigated. The measured energy response of LiF:Mg,Cu,P TLDs was fit to a simple model devised by Kron et al (1998 Phys. Med. Biol. 43 3235-59) to describe the variation of TLD response with x-ray energy. If TLDs are handled as recommended in the present paper, the energy response of LiF:Mg,Cu,P deviates by a maximum of 15% from unity and agrees with the model to within 5% or experimental uncertainty between 15 keV and 10 MeV. LiF:Mg,Cu,P TLDs of all forms have consistent and superior energy response compared to the standard material LiF:Mg,Ti and are therefore suitable for a wide range of applications in diagnostic radiology and radiotherapy.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/49/17/001DOI Listing

Publication Analysis

Top Keywords

lifmgcup tlds
16
response x-ray
12
x-ray energy
12
response lifmgcup
12
energy response
12
tlds forms
8
response
7
energy
6
lifmgcup
6
x-ray
5

Similar Publications

Background: A passive dosimeter framework for the measurement of dose in carbon ion beams has yet to be characterized or implemented for regular use.

Purpose: This work determined the dose calculation correction factors for absorbed dose in thermoluminescent dosimeters (TLDs) in a therapeutic carbon ion beam. TLD could be a useful tool for remote audits, particularly in the context of clinical trials as new protocols are developed for carbon ion radiotherapy.

View Article and Find Full Text PDF

Purpose: In vivo dosimetry is a common requirement to validate dose accuracy/uniformity in total body irradiation (TBI). Several detectors can be used for in vivo dosimetry, including thermoluminescent dosimeters (TLDs), diodes, ion chambers, optically stimulated luminescent dosimeters (OSLDs), and film. TLDs are well established for use in vivo but required expertise and clinical system availability may make them impractical for multifractionated TBI.

View Article and Find Full Text PDF

This research reviews a novel artificial intelligence (AI)-based application called TLDetect, which filters and classifies anomalous glow curves (GCs) of thermoluminescent dosimeters (TLDs). Until recently, GC review and correction in the lab were performed using an old in-house software, which uses the Microsoft Access database and allows the laboratory technician to manually review and correct almost all GCs without any filtering. The newly developed application TLDetect uses a modern SQL database and filters out only the necessary GCs for technician review.

View Article and Find Full Text PDF

Purpose: To compare the effect of two linacs designs on fetal dose sparing on a pregnant patient, including estimation of the fetal dose, and the effect of a lead apron.

Methods: A patient with a high-grade sarcoma located in the right knee/lower thigh was prescribed 51 Gy (1.7 Gy/Fx) with a simultaneous-integrated-boost (SIB) of 60 Gy to a smaller volume, starting in the 26th gestational week.

View Article and Find Full Text PDF

Introduction: Narrowed tracheal lumen diameter (TLD) in dogs caused by congenital hypoplasia or acquired tracheal stenosis can result in adverse health effects. Standardized tracheal scores calculated from radiographic measurements have been used to assess tracheal diameter however comparisons have not been made to characterize differences in tracheal lumen among breeds.

Methods: The main objective of this study was to compare tracheal scores at three regions of the trachea among non-brachycephalic dogs, non-bulldog brachycephalic dogs, and bulldogs.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!