Background: Carotid endarterectomy is more effective than medical management in the prevention of stroke in patients with severe symptomatic or asymptomatic atherosclerotic carotid-artery stenosis. Stenting with the use of an emboli-protection device is a less invasive revascularization strategy than endarterectomy in carotid-artery disease.

Methods: We conducted a randomized trial comparing carotid-artery stenting with the use of an emboli-protection device to endarterectomy in 334 patients with coexisting conditions that potentially increased the risk posed by endarterectomy and who had either a symptomatic carotid-artery stenosis of at least 50 percent of the luminal diameter or an asymptomatic stenosis of at least 80 percent. The primary end point of the study was the cumulative incidence of a major cardiovascular event at 1 year--a composite of death, stroke, or myocardial infarction within 30 days after the intervention or death or ipsilateral stroke between 31 days and 1 year. The study was designed to test the hypothesis that the less invasive strategy, stenting, was not inferior to endarterectomy.

Results: The primary end point occurred in 20 patients randomly assigned to undergo carotid-artery stenting with an emboli-protection device (cumulative incidence, 12.2 percent) and in 32 patients randomly assigned to undergo endarterectomy (cumulative incidence, 20.1 percent; absolute difference, -7.9 percentage points; 95 percent confidence interval, -16.4 to 0.7 percentage points; P=0.004 for noninferiority, and P=0.053 for superiority). At one year, carotid revascularization was repeated in fewer patients who had received stents than in those who had undergone endarterectomy (cumulative incidence, 0.6 percent vs. 4.3 percent; P=0.04).

Conclusions: Among patients with severe carotid-artery stenosis and coexisting conditions, carotid stenting with the use of an emboli-protection device is not inferior to carotid endarterectomy.

Download full-text PDF

Source
http://dx.doi.org/10.1056/NEJMoa040127DOI Listing

Publication Analysis

Top Keywords

stenting emboli-protection
16
emboli-protection device
16
cumulative incidence
16
carotid-artery stenting
12
carotid-artery stenosis
12
endarterectomy
8
carotid endarterectomy
8
patients severe
8
coexisting conditions
8
stenosis percent
8

Similar Publications

Pre-navigation balloon technique: Distal emboli protection during stent retriever thrombectomy.

Clin Neurol Neurosurg

January 2024

Department of Radiology, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, South Korea. Electronic address:

Objective: During stent retriever thrombectomy, a balloon guide catheter reduces distal emboli and consequently improves clinical outcomes. Because balloons are usually used before stent retrieval, these can affect the thrombus including the distal emboli while performing microcatheter navigation. This study aimed to evaluate the usefulness and safety of the pre-navigation balloon technique during microcatheter and microwire navigation.

View Article and Find Full Text PDF

Endovascular intervention has become the mainstay of treatment for subclavian artery stenosis in many centers, with high technical success and low complication rates.1,2 However, potential embolization during proximal subclavian artery intervention can lead to catastrophic posterior circulation ischemic complications.3-5 Although considered a rare complication, the presence of a contralateral hypoplastic vertebral artery with persisting anterograde vertebral blood flow on the affected side is likely to increase the risk of embolization.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!