Linear ion traps in mass spectrometry.

Mass Spectrom Rev

Department of Chemistry, University of British Columbia, 2036 Main Mall, Vancouver, BC, V6T 1Z1, Canada.

Published: February 2005

Linear ion traps are finding new applications in many areas of mass spectrometry. In a linear ion trap, ions are confined radially by a two-dimensional (2D) radio frequency (RF) field, and axially by stopping potentials applied to end electrodes. This review focuses on linear ion trap instrumentation. Potentials and ion motion in linear multipole fields and methods of ion trapping, cooling, excitation, and isolation are described. This is followed by a description of various mass discrimination effects that have been reported with linear ion traps. Linear ion traps combined in various ways with three-dimensional (3D) traps, time-of-flight (TOF) mass analyzers, and Fourier transform ion cyclotron resonance mass spectrometers are then given. Linear ion traps can be used as stand alone mass analyzers, and their use for mass analysis by Fourier transforming image currents, by mass selective radial ejection, and by mass selective axial ejection are reviewed.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mas.20004DOI Listing