Right-sided circulatory failure (RSCF) is a serious complication in 15-30% of patients receiving a left ventricular assist device (LVAD). It is hypothesized that left ventricular support which lacks physiologic properties predisposes to RSCF. An integral computer simulation and experimental validation protocol was performed. The results suggest that with conventional insensitive left ventricular support right-sided circulatory function is compromised, which may form a substrate for the onset or progress of RSCF. Feedback control of the LVAD could provide a means to counter this problem. A control concept for the LVAD which aims to preserve right-sided circulatory function, while supporting peripheral perfusion, is proposed

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1525-1594.2004.07380.xDOI Listing

Publication Analysis

Top Keywords

left ventricular
16
right-sided circulatory
16
ventricular assist
8
circulatory failure
8
ventricular support
8
circulatory function
8
physiologic-insensitive left
4
ventricular
4
assist predisposes
4
right-sided
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?