Purpose: The aim of this study was to evaluate the threshold levels of aberration change that a typical reference eye is able to detect.

Methods: The method involved simulation of the foveal vision of a typical eye in polychromatic light through optics affected by different levels of the various chosen monochromatic aberrations. The reference eye had the following monochromatic wavefront characteristics based on the aberrations of a population of young adults: no spherical defocus, astigmatism -0.37D oriented at 0 degrees celsius, coma -0.17 D/mm oriented at 270 degrees celsius, and spherical aberration -0.12 D/mm. Average amounts of longitudinal and transverse chromatic aberration were assumed, and allowance was made for the Stiles-Crawford effect. The pupil diameter of the simulated eye was kept fixed at 6 mm. Three observers each compared, 100 times, a simulated image as seen through the standard reference eye with a variant "aberrated" image. The varying parameter was the value of a chosen additional aberration affecting the variant image in the reference eye. The test was repeated for varying amounts of spherical defocus, astigmatic defocus, and spherical aberration. For each of these aberrations and each observer, the discrimination probability as a function of the aberration level in the variant image was determined. The just-noticeable difference in aberration (JNDA) was derived from each discrimination curve as the difference between the aberrations corresponding to discrimination probabilities of 75% and 25%. The JNDA values obtained were expressed in the form of root mean square (RMS) wavefront error thresholds.

Results: It was found that 0.04 microm of RMS aberration should be considered as the threshold of just-noticeable image change, in good agreement with the Maréchal criterion.

Conclusions: The results imply that in normal viewing conditions (e.g., a 3-mm pupil size), optical corrections should be in the range of +/-0.15 D in sphere and cylinder from the target prescription if perceptible change in the quality of the perceived images is to be avoided. The design of conventional soft contact lenses of high negative power or positive power should aim to produce -0.07 D/mm of spherical aberration, with a tolerated interval between -0.15 to +0.01 D/mm for a 6-mm pupil size.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.opx.0000144751.11213.cdDOI Listing

Publication Analysis

Top Keywords

spherical aberration
16
reference eye
16
aberration
10
spherical defocus
8
degrees celsius
8
variant image
8
pupil size
8
spherical
6
eye
6
image
5

Similar Publications

Clinical Relevance: Horizontal rectus muscle surgeries may cause changes in corneal and anterior segment parameters. Corneal topography is an important device for identifying these alterations.

Background: The aim of this work is to evaluate the effects of horizontal rectus muscle surgeries on corneal topography, anterior chamber parameters and corneal wavefront aberrations.

View Article and Find Full Text PDF

A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses.

J Clin Med

December 2024

The David J Apple Center for Vision Research, Department of Ophthalmology, Heidelberg University Eye Clinic, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany.

This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions. With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea.

View Article and Find Full Text PDF

We aimed to synthesize silver nanoparticles (AgNPs) using (cardamom) extracts and assess the cytotoxicity and genotoxicity of the cardamom extract, -AgNPs, and the insecticide ATCBRA-commonly used for pest control-on the root system of (broad bean). The chemical composition of the aqueous cardamom extract was identified and quantified using GC-MS, revealing a variety of bioactive compounds also present in cardamom essential oil. These included α-terpinyl acetate (21.

View Article and Find Full Text PDF

Purpose: To evaluate the clinical outcome of laser-assisted surgical correction of high hyperopic or mixed astigmatism using small incision intrastromal lenticule rotation (SMILERO) alone or combined with photorefractive keratectomy (PRK).

Methods: This retrospective case series enrolled 25 eyes with high astigmatism that underwent SMILERO surgery. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), uncorrected near visual acuity (UNVA), manifest refraction, central corneal thickness, and corneal higher order aberrations were analyzed before surgery and after 3, 6, and 12 months of follow-up.

View Article and Find Full Text PDF

Purpose: To compare the clinical outcomes, surgical workflow, and patient satisfaction following small incision lenticule extraction (SMILE) performed with the VisuMax 800 in one eye and the VisuMax 500 in the contralateral eye (both Carl Zeiss Meditec).

Methods: This was a prospective, single-site clinical study of patients undergoing SMILE for myopia and myopic astigmatism between February 2022 and August 2023. Each patient underwent bilateral treatment using the VisuMax 800 (VM800 group) in one eye and the VisuMax 500 (VM500 group) in the contralateral eye.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!