Objective: Hemodynamic performances of mechanical valve prostheses in patients with aortic valve stenosis and a small aortic annulus are crucial. We analyzed the in vitro hydrodynamics of 5 currently available bileaflet mechanical prostheses that fitted a 21-mm-diameter valve holder of a Sheffield pulse duplicator.

Methods: Three samples of 5 high-performance production-quality prostheses, including the sewing ring cuffs, were tested in the aortic chamber of a Sheffield pulse duplicator. Sizes of the prostheses fitting the 21-mm valve holder were as follows: 18-mm ATS, 19-mm SJM Regent, 19-mm Sorin Bicarbon Slimline, 19-mm On-X, and 21-mm Carbomedics Top Hat. The tests were carried out at a fixed pulse rate (70 beats/min) and at increasing cardiac outputs of 2, 4, 5, and 7 L/min. Each valve was tested 10 times for each different cardiac output. This resulted in a total of 40 tests for each valve and 120 tests for each valve model. The aortic pressure was set at 120/80 mm Hg (mean pressure, 100 mm Hg) throughout the experiment for all cardiac outputs. Forward flow pressure decrease, total regurgitant volume, closing and leakage volumes, effective orifice area, and stroke work loss were recorded while the valve operated under each cardiac output.

Results: The SJM Regent valve and the Sorin Bicarbon Slimline prosthesis showed the lowest mean and peak gradients at increasing cardiac outputs. The closure volume was higher for the SJM Regent and Sorin Bicarbon Slimline prostheses, unlike with the ATS prosthesis at 7 L/min of cardiac output. The ATS and SJM Regent prostheses showed the largest regurgitant volume, whereas the Sorin Bicarbon Slimline prosthesis showed the lowest regurgitant volume. The calculated effective orifice area and stroke work loss were significantly better with the SJM Regent and Sorin Bicarbon Slimline prostheses.

Conclusion: Assuming that the 21-mm valve holder in which all the tested prostheses were accommodated is comparable with a defined aortic valve size, this hydrodynamic evaluation model allowed us to compare the efficiency of currently available valve prostheses, and among these, the SJM Regent and the Sorin Bicarbon Slimline exhibited the best performances.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jtcvs.2004.03.021DOI Listing

Publication Analysis

Top Keywords

sjm regent
24
sorin bicarbon
24
bicarbon slimline
24
valve
12
valve holder
12
cardiac outputs
12
regurgitant volume
12
regent sorin
12
small aortic
8
aortic annulus
8

Similar Publications

New insights and novel perspectives in bileaflet mechanical heart valve prostheses thromboresistance.

J Cardiothorac Surg

May 2024

Department of Cardiothoracic Surgery, Heart Centre, University of Groningen, University Medical Centre Groningen, P.O. Box 30.001, Groningen, 9700 RB, The Netherlands.

Background: Although well-known for their thromboresistance, bileaflet mechanical heart valves (BMHV) require lifelong anti-thrombotic therapy. This must be associated with a certain level of thrombogenicity. Since both thromboresistance and thrombogenicity are explained by the blood-artificial surface or liquid-solid interactions, the aim of the present study was to explore BMHV thromboresistance from new perspectives.

View Article and Find Full Text PDF

Neoantigens are peptides derived from non-synonymous mutations presented by human leukocyte antigens (HLAs), which are recognized by antitumour T cells. The large HLA allele diversity and limiting clinical samples have restricted the study of the landscape of neoantigen-targeted T cell responses in patients over their treatment course. Here we applied recently developed technologies to capture neoantigen-specific T cells from blood and tumours from patients with metastatic melanoma with or without response to anti-programmed death receptor 1 (PD-1) immunotherapy.

View Article and Find Full Text PDF
Article Synopsis
  • Splenic abscess, a serious complication of infective endocarditis, was identified in a 49-year-old man treated with antibiotics after blood cultures showed Group G Streptococcus.* -
  • Imaging revealed a splenic abscess, leading to percutaneous drainage, and an echocardiogram showed a mobile vegetation on the aortic valve, which resulted in an aortic valve replacement.* -
  • The patient had successful outcomes with no infection recurrence 13 months later, highlighting a debate on whether to perform splenectomy first or valve replacement, given the embolism risk.*
View Article and Find Full Text PDF

Possible Early Generation of Physiological Helical Flow Could Benefit the Triflo Trileaflet Heart Valve Prosthesis Compared to Bileaflet Valves.

Bioengineering (Basel)

December 2020

School of Mathematics, Computer Science and Engineering, City, University of London, Northampton Square, London EC1V 0HB, UK.

Background-Physiological helical flow in the ascending aorta has been well documented in the last two decades, accompanied by discussions on possible physiological benefits of such axial swirl. Recent 4D-MRI studies on healthy volunteers have found indications of early generation of helical flow, early in the systole and close to the valve plane. Objectives-Firstly, the aim of the study is to investigate the hypothesis of premature swirl existence in the ventricular outflow tract leading to helical flow in the valve plane, and second to investigate the possible impact of two different mechanical valve designs on the preservation of this early helical flow and its subsequent hemodynamic consequences.

View Article and Find Full Text PDF

Comparative Study of Wall-Shear Stress at the Ascending Aorta for Different Mechanical Heart Valve Prostheses.

J Biomech Eng

January 2020

School of Mathematics, Computer Science and Engineering, City, University of London, Northampton Square, London EC1V 0HB, UK.

An experimental study is reported which investigates the wall shear stress (WSS) distribution in a transparent model of the human aorta comparing an St. Jude Medical (SJM) Regent bileaflet mechanical heart valve (BMHV) with the Lapeyre-Triflo FURTIVA trileaflet mechanical heart valve (TMHV) in physiological pulsatile flow. Elastic microcantilever structures, calibrated as micropillar WSS sensors by microparticle-image-velocimetry measurements, are applied to the wall along the ascending aorta (AAo).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!