A novel flow-injection method (FIA) for the determination of dopamine based on the inhibition of the intensity of chemiluminescence (CL) from luminol-hexacyanoferrate(III) system in basic medium is described. The present method allows the determination of dopamine over the range 30-100 microg l(-1) and 400-3000 microg l(-1). The relative standard deviation is 2.32% for 70 microg l(-1) dopamine and 1.22% for 1500 microg l(-1) dopamine (n = 20). The detection limit is 5 microg l(-1) with the sampling rate of 135 samples h(-1). This method has been applied for the determination of dopamine in commercial pharmaceutical injection samples. The results obtained by this method agreed with those by the official method.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jpba.2004.05.009DOI Listing

Publication Analysis

Top Keywords

microg l-1
20
determination dopamine
16
l-1 dopamine
8
method
5
dopamine
5
microg
5
l-1
5
determination
4
dopamine flow-injection
4
flow-injection analysis
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!