The optimum dose of gamma radiation-heavy doses to low wear polyethylene in total hip prostheses.

J Mater Sci Mater Med

Department of Orthopaedic Surgery, Artificial Joint Section and Biomaterial Research Laboratory, Osaka-Minami National Hospital, 677-2, Kido-Cho, Kawachinagano-Shi, Japan.

Published: January 1997

Wear volume, surface area and coefficient of friction of UHMWPE cup crosslinked with gamma radiation of 0, 50, 75, 100, 125, 150 and 200 Mrad sliding against an alumina ball were measured using a sphere-on-flat reciprocating type tribology testing machine. The effects of gamma radiation were scarcely observed in coefficient of friction. The coefficient of friction under lubricated (distilled water) and non-lubricated testing conditions was 0.08 to 0.12 and 0.20 to 0.25, respectively. The wear volume of UHMWPE with radiation of 50 Mrad, 75 to 150 Mrad, and 200 Mrad was 70 to 80%, 18 to 25%, and 12 to 15%, respectively, in comparison to non-irradiated specimens. Elongation and tensile strength of UHMWPE with radiation of 100 Mrad decreased to 6%, and 50% of that without radiation, respectively. The hardness increased with increase of the radiation dose. From several kinds of tribological findings, mechanical strength tests, and studies of long-term clinical findings, it is concluded that approximately 200 Mrad is the optimum dose of gamma radiation for clinical use in total hip prostheses.

Download full-text PDF

Source
http://dx.doi.org/10.1023/a:1018582027349DOI Listing

Publication Analysis

Top Keywords

coefficient friction
12
gamma radiation
12
200 mrad
12
optimum dose
8
dose gamma
8
total hip
8
hip prostheses
8
wear volume
8
radiation 100
8
uhmwpe radiation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!