Glucose kinetics were investigated in subcutaneous tissue of rabbits, in which a percutaneous device was implanted. The device was used for collection of tissue fluid and as carrier of an amperometric glucose sensor. Changes in glycaemia were reflected in subcutaneous tissue fluid. However, a limited number of responses of the implanted sensors were observed. Histologic evaluation showed thin fibrous capsules surrounding the implants. Accumulations of inflammatory cells were observed inside the subcutaneous chamber. The experiments again showed that changes in blood glucose concentration can be measured in subcutaneous tissue fluid collected with a percutaneous device. Nevertheless, implanted glucose sensors could not reliably monitor these changes. Supported by our histological observations and sufficient in vitro performance, we suppose that the cellular reaction to the sensor plays an important role in this poor in vivo performance. In combination with adsorption of tissue fluid proteins, this results in a reversible deactivation of implanted sensors. The exact mechanisms involved in this process are currently unknown and need further investigation.