Power injection of contrast media via peripherally inserted central catheters for CT.

J Vasc Interv Radiol

Department of Radiology, St. Luke's Medical Center, 2900 West Oklahoma Avenue, Milwaukee, Wisconsin 53715, USA.

Published: August 2004

Purpose: To evaluate patient safety, catheter rupture rates, and computed tomography (CT) image quality when using peripherally inserted central catheters (PICCs) in vivo for the power injection of CT contrast media at standard injection rates.

Materials And Methods: A prospective study to evaluate the safety and effectiveness of power injection of contrast media via indwelling PICCs was performed. Single-lumen and double-lumen polyurethane PICCs (5 F) were injected in vivo with contrast media for clinical CT examinations at injection rates ranging from 1 mL/sec to 4 mL/sec. Data collected included PICC rupture rate, patient complications, injection rate, peak injection pressure, PICC length, PICC age, and quality of contrast enhancement on the CT images.

Results: One hundred ten power injections of PICCs for CT examinations were performed. There were 12 injections of single-lumen PICCs and 98 injections of double-lumen PICCs. The most common injection rate was 2 mL/sec, accounting for 89 of the 110 injections (81%). Two PICCs ruptured during power injection, both as a result of operator error. One of the PICCs that ruptured was clamped at the time of injection and the other one was kinked at its venous entry site. One additional PICC showed evidence of dysfunction; it ballooned without actually rupturing. No significant patient complications occurred. Contrast enhancement of the CT images was subjectively rated as average or above average in 95% of cases.

Conclusions: Contrast media can be power-injected via PICCs for routine CT examinations at a rate of 2 mL/sec, yielding satisfactory image quality without exposing patients to significant additional risk. Power injection rates greater than 2 mL/sec, as are typically used in CT angiography applications, were not fully evaluated by this study.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.RVI.0000128812.20864.ECDOI Listing

Publication Analysis

Top Keywords

power injection
20
contrast media
20
injection contrast
12
injection
10
piccs
9
peripherally inserted
8
inserted central
8
central catheters
8
image quality
8
injection rates
8

Similar Publications

Objectives: This work highlights the methods used to develop a multi-pulse 1726 nm laser system combined with bulk air-cooling for selective sebaceous gland (SG) photothermolysis using thermal imaging and software algorithms. This approach enables treating to a desired tissue temperature and depth to provide a safe, effective, reproducible, and durable treatment of acne.

Methods: We designed and built a 1726 nm laser system with a 40 W maximum power output, a highly controlled air-cooling device, and a thermal camera in the handpiece, which permits real-time temperature monitoring of the epidermis.

View Article and Find Full Text PDF

Cannabis inflorescences represent an important source of many high-value bioactive specialized metabolites, among which the family of terpenes or terpenoids that are the largest classes of natural products known. Besides their biological activities either alone or synergistic with other terpenoids and/or cannabinoids, they are responsible for their distinctive flavour. In this study, we exploited the separation power and identification capabilities of comprehensive two-dimensional gas chromatography coupled to mass spectrometry (GC×GC-MS) for the profiling of terpenes and terpenoids in cannabis inflorescences.

View Article and Find Full Text PDF

Optimizing energy resources is a major priority these days. Increasing household energy demand often leads to the deterioration of poorly sized distribution networks. This paper presents a method for energy compensation and optimization in radial distribution systems (ORDS).

View Article and Find Full Text PDF

Our extensive basic research on photodynamic therapy (PDT) application in models of intracranial malignant astrocytoma led to its clinical application for intracranial malignant astrocytoma in Japan. Having considered the safety and effectiveness of this pathology, we initiate a first-in-human clinical study of PDT for spinal cord malignant astrocytoma. This study has an open-label, single-arm design.

View Article and Find Full Text PDF

Unlabelled: Accurate estimation of the Lung Shunt Fraction (LSF) is a standard of care in yttrium-90 ( Y) radioembolization treatment planning to prevent excessive lung irradiation due to arterio-venous shunting in the liver. LSF is assessed using Tc macroaggregated albumin ( Tc-MAA) imaging, but this approach adds risk, complexity, and expense to the treatment planning. This study investigates the potential of Contrast-Enhanced Computed Tomography (CECT) as a non-invasive alternative for LSF estimation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!