A compact MRI system for measuring trabecular bone volume fraction (TBVF) of the calcaneus was developed with the use of a 0.21 T permanent magnet and portable MRI console. The entire system weighed < 600 kg and was installed in a 2 m x 2 m space. Two cross-sectional spin-echo images of a heel acquired with external reference phantoms (total measurement time = 5 min) were used to quantify the TBVF of the calcaneus. The linearity and reproducibility of the measurements were evaluated by means of proton density-adjusted phantoms. Comparative measurements with quantitative ultrasound (QUS) in groups of healthy female volunteers showed a relatively high positive correlation (R(2) = 0.4539, 0.2693) between TBVF and the speed of sound (SOS). These results demonstrate the potential of this new system for measuring bone density. Magn Reson Med 52:440-444, 2004.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mrm.20135 | DOI Listing |