Objective: The objective of this study was to introduce continuous glucose-error grid analysis (CG-EGA) as a method of evaluating the accuracy of continuous glucose-monitoring sensors in terms of both accurate blood glucose (BG) values and accurate direction and rate of BG fluctuations and to illustrate the application of CG-EGA with data from the TheraSense Freestyle Navigator.

Research Design And Methods: We approach the design of CG-EGA from the understanding that continuous glucose sensors (CGSs) allow the observation of BG fluctuations as a process in time. We account for specifics of process characterization (location, speed, and direction) and for biological limitations of the observed processes (time lags associated with interstitial sensors). CG-EGA includes two interacting components: 1) point-error grid analysis (P-EGA) evaluates the sensor's accuracy in terms of correct presentation of BG values and 2) rate-error grid analysis (R-EGA) assesses the sensor's ability to capture the direction and rate of BG fluctuations.

Results: CG-EGA revealed that the accuracy of the Navigator, measured as a percentage of accurate readings plus benign errors, was significantly different at hypoglycemia (73.5%), euglycemia (99%), and hyperglycemia (95.4%). Failure to detect hypoglycemia was the most common error. The point accuracy of the Navigator was relatively stable over a wide range of BG rates of change, and its rate accuracy decreased significantly at high BG levels.

Conclusions: Traditional self-monitoring of BG device evaluation methods fail to capture the important temporal characteristics of the continuous glucose-monitoring process. CG-EGA addresses this problem, thus providing a comprehensive assessment of sensor accuracy that appears to be a useful adjunct to other CGS performance measures.

Download full-text PDF

Source
http://dx.doi.org/10.2337/diacare.27.8.1922DOI Listing

Publication Analysis

Top Keywords

grid analysis
16
continuous glucose-monitoring
12
evaluating accuracy
8
accuracy continuous
8
glucose-monitoring sensors
8
continuous glucose-error
8
glucose-error grid
8
therasense freestyle
8
direction rate
8
accuracy navigator
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!