Objective: To describe a case of artifactually decreased hemoglobin Alc (HbAlc) attributable to use of dapsone.
Methods: We present a detailed case report and results of a related literature search. In addition, potential causes of artifactually lowered HbAlc values are discussed.
Results: A 35-year-old patient with type I diabetes had high home-monitored blood glucose values, high clinic plasma glucose determinations, increased fructosamine levels, and low HbAlc values. The lowering of the HbAlc level was associated with use of dapsone, and the decrease in HbAlc value was proportional to the dose of dapsone. A literature search revealed one previous report of artifactual lowering of the HbAlc value, in which high methemoglobin levels were found and thought to be the cause of the artifactually decreased HbA1c.
Conclusion: By increasing methemoglobin levels and decreasing erythrocyte survival, dapsone can artifactually lower HbA1c. Clinicians should be aware of this potential side effect of dapsone.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4158/EP.8.5.370 | DOI Listing |
Membranes (Basel)
October 2024
Department of Medical Physics and Biophysics, University of Split School of Medicine, 21000 Split, Croatia.
Giant unilamellar vesicles (GUVs) are frequently used as membrane models in studies of membrane properties. They are most often produced using the electroformation method. However, there are a number of parameters that can influence the success of the procedure.
View Article and Find Full Text PDFAdv Lab Med
September 2024
Clinical Laboratory, Division of Nephrology and Hypertension, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
Environ Sci Technol
August 2024
Department of Environmental Sciences, University of Basel, Klingelbergstrasse 27, 4056 Basel, Switzerland.
Secondary organic aerosol (SOA) represents a large fraction of atmospheric aerosol particles that significantly affect both the Earth's climate and human health. Laboratory-generated SOA or ambient particles are routinely collected on filters for a detailed chemical analysis. Such filter sampling is prone to artifactual changes in composition during collection, storage, sample workup, and analysis.
View Article and Find Full Text PDFSci Rep
June 2024
Biopsychology and Neuroergonomics, Institute of Psychology and Ergonomics, TU Berlin, Berlin, Germany.
Electroencephalography (EEG) studies increasingly utilize more mobile experimental protocols, leading to more and stronger artifacts in the recorded data. Independent Component Analysis (ICA) is commonly used to remove these artifacts. It is standard practice to remove artifactual samples before ICA to improve the decomposition, for example using automatic tools such as the sample rejection option of the AMICA algorithm.
View Article and Find Full Text PDFJ Am Coll Radiol
March 2024
Senior Vice President of Early Cancer Detection Science, American Cancer Society.
Purpose: Cancer detection rate (CDR), an important metric in the mammography screening audit, is designed to ensure adequate sensitivity. Most practices use biopsy results as the reference standard; however, commonly ascertainment of biopsy results is incomplete. We used simulation to determine the relationship between the cancer ascertainment rate of biopsy (AR-biopsy), CDR estimation, and associated error rates in classifying whether practices and radiologists meet the established ACR benchmark of 2.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!