AI Article Synopsis

  • The study compared the rotatory stability of hip endoprostheses implanted by robots versus manual methods.
  • The research involved various prosthesis-stems and assessed their stability after implantation in polyurethane blocks subjected to rotatory stress.
  • Results showed that robotic implantation generally improved stability, with the CBC stem performing best, while prosthesis design had a significant impact on stability regardless of the implantation technique.

Article Abstract

Aim: In our study we compared the primary rotatory stability of hip endoprostheses which were implanted with the help of a robot with that of manually implanted stems.

Method: We examined three different types of prosthesis-stems: Osteolock (Stryker-Howmedica), CBC (Mathys), Excia (Aesculap). Furthermore, we examined two types of prostheses which could not be implanted with a robot (Sl-Plus/Endoplus; Hipstar/Stryker-Howmedica). 10 stems of each prosthesis type were implanted in identical polyurethane foam blocks; 5 of them were implantated manually, 5 with the help of a robot (CASPAR, URS-Ortho). After implantation, a defined rotatory stress was put on the stem with the help of a torquing machine. The torsional moment which was necessary until the stem broke out of the foam was documented with the help of special software.

Results: In all types of prostheses the use of the robot system led to a higher rotatory stability compared to the manual group. Among the three tested types of prosthesis stems, the CBC stem is significantly more stable for rotatory forces after robot-assisted implantation compared to the other two types. After manual implantation there was no difference in stability between the different stems. The Zweymüller stem and the Hipstar prosthesis did very well, even in comparison to the robot-implanted group. The results of the Zweymüller stem and the robot-implanted CBC Stem were similar, and a significantly higher rotatory stability of the Hipstar prosthesis compared to every other stem was observed.

Conclusion: Using a robot, the rotatory stability of hip endoprostheses can be improved. However, the design of the stems seems to be even more important for the stability than the implantation technique.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2004-822666DOI Listing

Publication Analysis

Top Keywords

rotatory stability
20
stability hip
12
hip endoprostheses
12
help robot
8
types prostheses
8
higher rotatory
8
cbc stem
8
zweymüller stem
8
hipstar prosthesis
8
stability
7

Similar Publications

The anterolateral ligament (ALL) is considered a secondary stabilizer of internal rotation and, due to proximity to the lateral collateral ligament (LCL), it may contribute to anterolateral rotatory stability. This study characterized the anatomy of the ALL and associated tissues of the anterolateral complex (ALC) to determine if structural and histological compensatory adaptations exist in patients without an ALL. Forty-nine cadaveric knees were dissected from distal-to-proximal using established landmarks with the aid of internal rotation stress to localize the ALL (if present), LCL, iliotibial band (ITB), and anterolateral capsule.

View Article and Find Full Text PDF

: The elbow joint is stabilized by complex interactions between bony structures and soft tissues, notably the lateral and medial collateral ligaments. Posterolateral rotatory instability (PLRI), a form of elbow instability, is challenging to diagnose due to overlapping symptoms with other conditions. The radiocapitellar line (RCL) is a radiographic tool for assessing humeroradial alignment and elbow stability, but its diagnostic accuracy remains debated.

View Article and Find Full Text PDF

The lateral collateral ligament complex of the elbow is pivotal for maintaining the stability of the elbow joint. The open technique for reconstructing the lateral ulnar collateral ligament (LUCL) is a standard procedure to treat elbow instability caused by LUCL deficiency. Nevertheless, as arthroscopy procedures in the elbow have advanced, we describe an arthroscopic technique to reconstruct the LUCL with suture anchors and bone tunnel techniques.

View Article and Find Full Text PDF

Purpose: Varus posteromedial rotatory instability (VPMRI) involves anteromedial coronoid fracture (AMCF), lateral ulnar collateral ligament (LUCL), and medial collateral ligament (MCL) injury. There is no general consensus regarding the surgical treatment, but most surgeons recommend internal fixation of the coronoid along with primary ligament repair. This methodology involves postoperative immobilization to allow ligament healing, occasionally associated with stiffness.

View Article and Find Full Text PDF

Purpose: To synthesize the current literature on posterolateral corner (PLC) reconstruction and repair, focusing on a comparison of surgical techniques and outcomes based on injury chronicity in the setting of sports-related ligament tears.

Methods: A systematic review of the literature, including queries of the PubMed/MEDLINE, Embase, and Cochrane Library databases, was performed in March 2024. Studies were included if they were of Level IV evidence or greater, reported PLC reconstruction or repair outcomes, and had a minimum 2-year follow-up.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!