Background: The effective orifice area (EOA) is the standard parameter for the clinical assessment of aortic stenosis severity. It has been reported that EOA measured by Doppler echocardiography does not necessarily provide an accurate estimate of the cross-sectional area of the flow jet at the vena contracta, especially at low flow rates. The objective of this study was to test the validity of the Doppler-derived EOA.
Methods: Triangular and circular orifice plates, funnels, and bioprosthetic valves were inserted into an in vitro aortic flow model and were studied under different physiologic flow rates corresponding to cardiac outputs varying from 1.5 to 7 L/min. For each experiment, the EOA was measured by Doppler and compared with the catheter-derived EOA and with the EOA derived from a theoretic formula. In bioprostheses, the geometric orifice area (GOA) was estimated from images acquired by high-speed video recording.
Results: There was no significant difference between the EOA derived from the 3 methods with the rigid orifices (Doppler vs catheter: y = 0.97x +0.18 mm(2), r(2) = 0.98; Doppler vs theory: y = 1.00x -3.60 mm(2), r(2) = 0.99). Doppler EOA was not significantly influenced by the flow rate in rigid orifices. As predicted by theory, the average contraction coefficient (EOA/GOA) was around 0.6 in the orifice plates and around 1.0 in the funnels. In the bioprosthetic valves, both EOA and GOA increased with increasing flow rate whereas contraction coefficient was almost constant with an average value of 0.99. There was also a very good concordance between EOA and GOA (y = 0.94x +0.05 mm(2), r(2) = 0.88).
Conclusions: In rigid aortic stenosis, the Doppler EOA is much less flow dependent than generally assumed. Indeed, it depends mainly on the GOA and the inflow shape (flat vs funnel-shaped) of the stenosis. The flow dependence of Doppler EOA observed in clinical studies is likely a result of a variation of the valve GOA or of the valve inflow shape and not an inherent flow dependence of the EOA derived by the continuity equation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.echo.2004.03.030 | DOI Listing |
Can J Cardiol
January 2025
Department of Cardiology, Istituto Auxologico Italiano, IRCCS, Milan, Italy; Department of Medicine and Surgery, University of Milano- Bicocca, Milan, Italy.
Background: In patients with moderate and severe secondary tricuspid regurgitation (STR), the effective regurgitant orifice area (EROA), corrected using the proximal isovelocity surface area (PISA) method for tricuspid valve leaflet tethering and low TR jet velocities, has an unclear threshold for identifying high-risk patients. This study aimed to establish a risk-based EROA cutoff and assess the impact of right ventricular (RV) remodeling on outcomes in low-risk STR patients according to EROA.
Methods: We included 513 consecutive outpatients (age 75±13 years, 47% male) with moderate and severe STR.
Comput Biol Med
January 2025
UCL Mechanical Engineering, University College London, UK; Ri.MED Foundation, Palermo, Italy; University of Palermo, Department of Engineering, Palermo, Italy. Electronic address:
Aortic valve replacements, both surgical and transcatheter, are nowadays widely employed treatments. Although clinically effective, these procedures are correlated with potentially severe clinical complications which can be associated with the non-physiological haemodynamics that they establish. In this work, the fluid dynamics changes produced by surgical and transcatheter aortic valve replacements are analysed and compared with an ideal healthy native valve configuration, employing advanced fluid-structure interaction (FSI) simulations.
View Article and Find Full Text PDFJ Mech Behav Biomed Mater
January 2025
School of Materials Science and Engineering, Colorado State University, 700 Meridian Ave, Fort Collins, 80523, CO, USA. Electronic address:
Hyaluronic acid-enhanced polyethylene polymeric TAVR shows excellent in vivo anti-calcific, anti-thrombotic, and in vitro hydrodynamic performance. However, during durability testing, impact wear and fatigue cause early valve failure. Heart valve durability can be improved by strengthening the leaflet with fiber reinforcement.
View Article and Find Full Text PDFSci Rep
January 2025
Key Laboratory of Cosmetic, China National Light Industry, Beijing Technology and Business University, Beijing, China.
Oily sensitive skin is complex and requires accurate identification and personalized care. However, the current classification method relies on subjective assessment. This study aimed to classify skin type and subtype using objective biophysical parameters to investigate differences in skin characteristics across anatomical and morphological regions.
View Article and Find Full Text PDFJTCVS Open
December 2024
Department of Cardiovascular Surgery, Jefferson Health, Philadelphia, Pa.
Objective: To compare outcomes of aortic valve replacement (AVR) in patients with pure aortic stenosis (Pure AS) and those with pure aortic regurgitation (Pure AR) or mixed AS and AR (MAVD) in the COMMENCE trial.
Methods: Of 689 patients who underwent AVR in the COMMENCE trial, patients with moderate or severe AR with or without AS (Pure AR + MAVD; n = 135) or Pure AS (n = 323) were included. Inverse probability of treatment weighting Kaplan-Meier survival curves were used for time-to-event endpoints, and longitudinal changes in hemodynamics were evaluated using mixed-effects models.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!