Purpose: For almost 3 years, Radiologists and Physicists from Padova and Ferrara Universities have collaborated together, with the aim of collecting and comparing experimental data useful to define the most significant parameters for quality controls in digital mammography. Successively, radiologists and physicists working in other sites where a digital mammography unit was installed joined the work-group.

Materials And Methods: In this study we report the results obtained from measurements of linearity, uniformity, short- and long-term reproducibility, AEC stability performed on 5 digital equipment by using a simple test object. X-ray beam quality and tube yield were preliminarily checked in such a way that possible uncertainties of digital system responses could be separated from those due to differences among X-ray tubes.

Results: Results showed that the equipment considered, comparable in terms of both beam quality (HVL) and tube yield, always displayed linear response and reproducibility errors lower than 5%. Uniformity was very good and the grey level compensation as a function of exposure parameters remained within 5%. Differences in choice of parameters by exposure control system (AOP) were emphasised, especially for crossing between track/filter combinations (from Mo/Mo to Mo/Rh and from Mo/Rh to Rh/Rh); those differences were attributed to the +/-2 mm tolerance of breast thickness measurement (mechanically obtained) greater than the AOP tolerance (+/-1 mm).

Discussion And Conclusions: Obtained results can be useful, as comparison and reference values, for users employing a digital mammography unit of the same kind reported in this paper. Moreover, the same results could be used as "orientation" also by other users having different digital mammography technologies, whose operation should be nevertheless specifically studied and understood in order to find the most useful parameters for quality controls. The acquired experience clearly showed us that years of investigations will be necessary in order to be able to write reliable protocols. This should induce people to contemplate the necessity of not improvising "theoretical" protocols, that are unreliable and dangerous for their negative clinical implications.

Download full-text PDF

Source

Publication Analysis

Top Keywords

digital mammography
20
radiologists physicists
8
parameters quality
8
quality controls
8
mammography unit
8
beam quality
8
tube yield
8
digital
7
quality
5
mammography quality
4

Similar Publications

Breast arterial calcification (BAC) is a common benign finding on a screening mammogram. Additionally, BAC is a type of medial calcification known as Mönckeberg medial calcific sclerosis, which differs from the intimal calcification seen in patients with coronary artery disease (CAD). Recently, BAC has appeared as a new cardiovascular risk stratification method.

View Article and Find Full Text PDF

Background The incidence of margin re-excision following breast conserving surgery (BCS) is a quality measure in the National Health Service. The threshold is less than 20% of all BCS procedures. Despite three decades of studies and a wealth of literature identifying multiple factors associated with increased risk for margin involvement, an accepted threshold rate affecting one in five procedures remains high.

View Article and Find Full Text PDF

Background: Breast cancer screening (BCS) inequities are evident at national and local levels, and many health systems want to address these inequities, but may lack data about contributing factors. The objective of this study was to inform health system interventions through an exploratory analysis of potential multilevel contributors to BCS inequities using health system data.

Methods: The authors conducted a cross-sectional analysis within a large academic health system including 19,774 individuals who identified as Black (n = 1445) or White (n = 18,329) race and were eligible for BCS.

View Article and Find Full Text PDF

External validation of a multivariable prediction model for positive resection margins in breast-conserving surgery.

BMC Res Notes

January 2025

Department of Surgery, Department of Clinical Sciences, Division of Surgery, Skåne University Hospital, Lund University, Lund, Sweden.

Objectives: Positive resection margins after breast-conserving surgery (BCS) most often demands a repeat surgery. To preoperatively identify patients at risk of positive margins, a multivariable model has been developed that predicts positive margins after BCS with a high accuracy. This study aimed to externally validate this prediction model to explore its generalizability and assess if additional preoperatively available variables can further improve its predictive accuracy.

View Article and Find Full Text PDF

[Analysis of Factors That Promote Awareness of Breast MRI Surveillance for Carriers of Hereditary Breast Cancer Risk Genes ( BRCA1/2)].

Nihon Hoshasen Gijutsu Gakkai Zasshi

January 2025

Department of Risk Analysis and Biodosimetry, Institute of Radiation Emergency Medicine, Hirosaki University.

Purpose: Hereditary breast and ovarian cancers (HBOC) carry a high risk of breast cancer, and detailed screening with contrast-enhanced breast MRI (breast MRI surveillance) is recommended. With the increase in the number of individuals diagnosed with HBOC, the demand for breast MRI surveillance is also rising. However, the current system is inadequate, with factors such as lack of knowledge and indifference among healthcare professionals, and insufficient understanding of breast MRI surveillance being cited.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!