Rationale And Objectives: This study evaluated different stent grafts by 3 T magnetic resonance angiography (MRA) with respect to lumen visibility, susceptibility-induced signal loss, and type of stent artifacts compared with 1.5 T MRA in a phantom model.
Methods: Six different stent-grafts (tube: n = 3, bifurcated: n = 3) were evaluated by 3 T and 1.5 T MRA using a tube phantom. MRA was performed using T1-weighted sequences at both systems with comparable parameters (3T: TR 5.4/TE 2.0/FA 30 degrees, 1.5 T: TR 6.2/TE 2.2/FA 30 degrees). A blind study of the image quality, including artifacts, was performed by 3 radiologists. Furthermore, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) values were calculated. Statistical analysis was performed with Student's t test (P < 0.05).
Results: One Elgiloy stent graft showed almost a complete intraluminal signal loss at 1.5 and 3 T. All other models could be evaluated by both systems by MRA, resulting in a favorable lumen visibility (score: 1) for prostheses made of nitinol. Scores for overall image quality and artifacts were the same for both MR systems. SNR and CNR values of the stented part of the vessel phantom increased from 320 +/- 33 to 618 +/- 40 and from 306 +/- 34 to 596 +/- 40 at 3 T when compared with 1.5 T, resulting in a significant signal gain of 93% at the higher field strength.
Conclusions: 3 Tesla MRA of aortic stent grafts in a phantom model demonstrates an increase in SNR and CNR when compared with 1.5 T. However, the magnitude of imaging artifacts as well as coherent intraluminal signal loss within the stent does not increase equally in both MR systems.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/01.rli.0000129470.63758.c1 | DOI Listing |
J Clin Med
January 2025
Department of Thoracic Surgery, Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo, 200, 00128 Rome, Italy.
Rigid bronchoscopy (RB) is the gold standard for managing central airway obstruction (CAO), a life-threatening condition caused by both malignant and benign etiologies. Anesthetic management is challenging as it requires balancing deep sedation with maintaining spontaneous breathing to avoid airway collapse. There is no consensus on the optimal anesthetic approach, with options including general anesthesia with neuromuscular blockers or spontaneous assisted ventilation (SAV).
View Article and Find Full Text PDFBiomedicines
January 2025
Department of Orthopaedic Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
Subclavian artery pseudoaneurysms are rare but potentially life-threatening vascular injuries frequently associated with trauma such as clavicle fractures. In this paper we describe the case of a 49-year-old male who developed a post-traumatic pseudoaneurysm of the subclavian artery after a bicycle accident. The diagnosis was delayed due to non-specific symptoms and an initially missed aneurysm on computed tomography imaging.
View Article and Find Full Text PDFDiagnostics (Basel)
January 2025
Vascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I-"La Sapienza" University of Rome, Viale del Policlinico, 00161 Rome, Italy.
: This study aimed to evaluate whether gender-related anatomical differences in proximal aortic neck morphology affect the feasibility and outcomes of endovascular aortic aneurysm repair (EVAR) in women with abdominal aortic aneurysms (AAA). : This study performed a retrospective analysis of patients electively treated by EVAR for infrarenal AAA between January 2019 and December 2023. Demographics, anatomical characteristics, and stent graft details were analyzed.
View Article and Find Full Text PDFJ Cardiovasc Dev Dis
January 2025
Department of Cardiac Surgery, Lankenau Heart Institute, Main Line Health, Wynnewood, PA 19096, USA.
Objectives: The impact of long-term complications after robotic hybrid coronary revascularization (HCR), including persistent angina, repeat revascularization, and myocardial infarction (MI), remains limited. This study aims to determine the risk factors for coronary events after robotic HCR and their time-varying effects on outcomes.
Methods: We identified all consecutive patients who underwent robotic HCR at our institution.
J Endovasc Ther
January 2025
Department of Vascular Surgery, The Chaim Sheba Medical Center, Tel HaShomer, Ramat Gan, Israel.
Purpose: To report a case series on using a novel semi-branch feature in custom-made stent-grafts in the endovascular treatment of complex aortic aneurysms and summarize the contemporary usage of this technology.
Case Series: Four patients underwent endovascular aortic aneurysm repair (EVAR) with a custom-made semi-branch stent-graft (Semi-Branch Endovascular Aortic Aneurysm Repair [SBEVAR]). Two male patients, 75- and 76-year-old, were treated due to failed EVAR with late-type Ia endoleak, and the other two, 80- and 55-year-old male patients, due to a juxta-renal aortic abdominal aneurysm (JRAAA).