To gauge the experimental variability associated with Biacore analysis, 36 different investigators analyzed a small molecule/enzyme interaction under similar conditions. Acetazolamide (222 g/mol) binding to carbonic anhydrase II (CAII; 30000 Da) was chosen as a model system. Both reagents were stable and their interaction posed a challenge to measure because of the low molecular weight of the analyte and the fast association rate constant. Each investigator created three different density surfaces of CAII and analyzed an identical dilution series of acetazolamide (ranging from 4.1 to 1000 nM). The greatest variability in the results was observed during the enzyme immobilization step since each investigator provided their own surface activating reagents. Variability in the quality of the acetazolamide binding responses was likely a product of how well the investigators' instruments had been maintained. To determine the reaction kinetics, the responses from the different density surfaces were fit globally to a 1:1 interaction model that included a term for mass transport. The averaged association and dissociation rate constants were 3.1+/-1.6 x 10(6)M(-1)s(-1) and 6.7+/-2.5 x 10(-2)s(-1), respectively, which corresponded to an average equilibrium dissociation constant (K(D) of 2.6+/-1.4 x 10(-8)M. The results provide a benchmark of variability in interpreting binding constants from the biosensor and highlight keys areas that should be considered when analyzing small molecule interactions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ab.2004.02.027DOI Listing

Publication Analysis

Top Keywords

small molecule/enzyme
8
molecule/enzyme interaction
8
density surfaces
8
comparative analyses
4
analyses small
4
interaction
4
interaction multiple
4
multiple users
4
users biacore
4
biacore technology
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!