Objective assessment of frequency-specific hearing thresholds in babies.

Int J Pediatr Otorhinolaryngol

Lab. Exp. ORL, K.U. Leuven, Kapucijnenvoer 33, B-3000 Leuven, Belgium.

Published: July 2004

Objective: To report on clinical experience using dichotic multiple-stimulus auditory steady-state responses (ASSRs) as an objective technique to estimate frequency-specific hearing thresholds in hearing-impaired infants.

Methods: A comparison was made between the click-evoked auditory brainstem response (ABR), auditory steady-state responses and behavioral hearing thresholds (BHTs). Both ears of 10 infants between 3 and 14 months of age were tested. ABR and ASSRs were recorded during the same test session. ABR was evoked by 100 micros clicks. ASSRs were evoked by amplitude- and frequency-modulated tones with carrier frequencies of 0.5, 1, 2 and 4 kHz and modulation frequencies ranging from 82 to 110 Hz. Eight signals (four to each ear) were presented simultaneously. ASSR thresholds were derived after separate recordings of approximately 5, 7.5 and 10 min to compare the influence of test duration. BHTs were defined in later test sessions as soon as possible after the ASSR test, dependent on medical and developmental factors.

Results: For the subjects tested in this study 60% of ABR thresholds and 95% of ASSR thresholds for 1, 2 and 4 kHz were found at an average age of 7 months. Only 51% of frequency-specific BHTs could be obtained but on average 5 months later. The correlation of ABR thresholds and ASSR thresholds at 2 kHz was 0.77. The correlation of ASSRs and BHTs was 0.92. The mean differences and associated standard deviations were 4 +/- 14, 4 +/- 11, -2 +/- 14 and -1 +/- 13 dB for 0.5, 1, 2 and 4 kHz, respectively. The average test duration was 45 min for ABR (one threshold in both ears) and 58 min for ASSR (four thresholds in both ears). By reducing the duration of the separate recordings of ASSR, the precision of the hearing threshold estimate decreased and the number of outlying and missing values increased. Correlation coefficients were 0.92, 0.89 and 0.83 for recordings of maximum 10, 7.5 and 5 min, respectively. A compromise between test duration and precision has to be sought.

Conclusions: Multiple-frequency ASSRs offer the possibility to estimate frequency-specific hearing thresholds in babies in a time-efficient way.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijporl.2004.02.007DOI Listing

Publication Analysis

Top Keywords

hearing thresholds
16
assr thresholds
16
frequency-specific hearing
12
test duration
12
+/- +/-
12
thresholds
10
thresholds babies
8
auditory steady-state
8
steady-state responses
8
estimate frequency-specific
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!