A theoretical analysis by a finite elements model (FEM) of some external fixators (Hoffmann, Wagner, Orthofix and Ilizarov) was carried out. This study considered a logarithmic progress of callus elastic characteristics. A standard configuration of each fixator was defined where design and application characteristics were modified. A comparison among standard configurations and influence of every variation was made with regard to displacement and load transmission at the fracture site. An experimental evaluation of standard configurations was performed with a testing machine. After experimental validation of the theoretical model was achieved, an application of physiological loads which act on a fractured limb during normal gait was analysed. A minimal contribution from an external fixator to the total rigidity of the bone-callus-fixator system was assessed when a callus showing minimum elastic characteristics had just been established. Insufficient rigidity from the fixation devices to assure an adequate immobilization during the early stages of fracture healing was verified. However, regardless of the external fixator, callus development was the overriding element for the rigidity of the fixator-bone system.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0021-9290(92)90035-yDOI Listing

Publication Analysis

Top Keywords

callus development
8
hoffmann wagner
8
wagner orthofix
8
orthofix ilizarov
8
external fixators
8
elastic characteristics
8
standard configurations
8
external fixator
8
biomechanical consequences
4
callus
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!