This study was designed to investigate the intratester and intertester reliability of isokinetic ankle inversion and eversion-strength measurement in neutral foot position in healthy adults using the Biodex dynamometer. Twenty-five men and women performed five maximal concentric contractions at 60 and 180 degrees/s angular velocities. Two physicians tested each subject. The first physician applied the test four times, and the second physician three times. Reliability of peak torque was assessed by calculating the intraclass correlation coefficient (ICC). At both angular velocities, inversion strength was greater than eversion, and when the angular velocity was increased, inversion and eversion strength were decreased, as tested by both physicians. The first measurements of inversion and eversion strength of the first physician were significantly lower than the other measurements (p<0.01). The intratester ICCs for ankle inversion in healthy young adults were highly reliable (ICC 0.92-0.96), and for the eversion values ranged from 0.87 to 0.94. Intertester ICCs for ankle inversion and eversion peak torque values demonstrated a value of 0.95. Isokinetic tests of ankle inversion and eversion strength at 60 and 180 degrees/s angular velocities in neutral foot position for healthy adults are highly reliable with the Biodex dynamometer.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00167-004-0530-8DOI Listing

Publication Analysis

Top Keywords

reliability isokinetic
8
isokinetic ankle
8
eversion-strength measurement
8
measurement neutral
8
neutral foot
8
foot position
8
biodex dynamometer
8
angular velocities
8
inversion eversion
8
eversion strength
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!