Rationale And Objectives: To determine which factors affected the increase in average glandular dose recorded at the annual US Food and Drug Administration Mammography Quality Standards Act inspections of mammography equipment in North Carolina from 1997 to 2001.
Materials And Methods: Average glandular dose, HVL, kVp, ambient light, luminance, equipment age, processing speed, and system speed for every mammography unit at all facilities in the state were collected by state inspectors. A mixed-effect model was used to assess the average changes of glandular dose over time and to identify the factors associated with these changes.
Results: There was a statistically significant increase in the average glandular dose in North Carolina in 1999, 2000, and 2001 when compared with the baseline year of 1997. Factors that were statistically significantly linked to this effect were changes in kVp, processing speed, and system speed.
Conclusion: Average glandular dose for mammography has recently increased in North Carolina. This change is likely caused by changes in screen-film products and processing techniques.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.acra.2004.01.003 | DOI Listing |
Phys Med
January 2025
Medical Physics Unit, ASST Monza, Monza, Italy.
Purpose: Digital Breast Tomosynthesis (DBT) is an advanced mammography technique for which there are currently no internationally agreed methods and reference values for image quality assessment. The aim of this multicentre study was to evaluate a simple method to assess the technical image quality of reconstructed and synthetic 2D (SM) images of different models of DBT systems using commercially available phantoms.
Methods: The signal difference to noise ratio (SDNR) was chosen as an index of technical image quality and was evaluated for three commercial phantoms, Tomophan, Tormam and CIRS model 015, on 55 DBT systems (six vendors, nine models).
Phys Med
January 2025
National Co-ordinating Centre for the Physics of Mammography (NCCPM), Medical Physics Department, Royal Surrey NHS Foundation Trust, Guildford GU2 7XX UK.
Background: The American Association of Physicists in Medicine (AAPM) Task group 282 (TG282) in collaboration with the European Federation for Organisations of Medical Physics (EFOMP) have developed a novel breast dosimetry model intended as a single international standard.
Purpose: To explore the impact of TG282 dosimetry on estimates of average Mean Glandular Dose (MGD) in the United Kingdom (UK) National Health Service (NHS) Breast Screening Programmes (BSP).
Methods: MGDs were estimated, using the TG282 dosimetry model, for the most recent UK NHSBSP dose survey.
Med Phys
January 2025
Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
Background: Dedicated breast computed tomography (bCT) systems offer detailed imaging for breast cancer diagnosis and treatment. As new bCT generations are developed, it is important to evaluate their imaging performance and dose efficiency to understand differences over previous models.
Purpose: To characterize the imaging performance and dose efficiency of a second-generation (GEN2) bCT system and compare them to those of a first-generation (GEN1) system.
AJR Am J Roentgenol
January 2025
Department of Radiology, Division of Breast Imaging and Intervention, Mayo Clinic, Phoenix, AZ.
Contrast-enhanced mammography (CEM) is growing in clinical use due to its increased sensitivity and specificity compared to full-field digital mammography (FFDM) and/or digital breast tomosynthesis (DBT), particularly in patients with dense breasts. To perform an intraindividual comparison of MGD between FFDM, DBT, a combination protocol using both FFDM and DBT (combined FFDM-DBT), and CEM, in patients undergoing breast cancer screening. This retrospective study included 389 women (median age, 57.
View Article and Find Full Text PDFCancer Manag Res
January 2025
School of Physics, Universiti Sains Malaysia, Penang, 11800 Gelugor, Malaysia.
Introduction: Breast cancer is a significant worldwide health issue, particularly in Jordan, where early detection via mammography is essential for effective disease management. Despite the little radiation risk associated with mammography, it is crucial to monitor radiation exposure to guarantee patient safety. This study intends to assess skin entrance exposure and compute the Mean Glandular Dose (MGD) in mammography units to determine adherence to established criteria and pinpoint areas for enhancement.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!