[Digital fluoroscopy imaging system with dual energy X-ray image decomposition].

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi

Institute of Biomedical Engineering, Xi'an Jiaotong University, Xi'an 710049, China.

Published: April 2004

AI Article Synopsis

Article Abstract

Dual-kilovolt technique was realized in dual energy X-ray imaging system with computer controlled KV adjustment, filter switching and image acquirement. The dual energy images can be obtained in 2-3 seconds to avoid moving artifact. Low energy photons in high energy X-ray beam can be attenuated effectively by a 2 mm Cu filter to separate the spectra of the high and low energy X-rays. A simplified dual energy decomposition method was used to obtain the soft tissue image. An example of chest dual energy radiographs is presented to illustrate the performance of the system.