Objective: To compare the reliability of central corneal thickness measurements (CCT) obtained with partial coherence interferometry (PCI), ultrasound pachymetry, and the Orbscan system.

Design: Cross-sectional study.

Participants: Twenty healthy subjects with CCT measurements in both eyes.

Methods: The CCT measurements were obtained with PCI, ultrasound pachymetry, and the Orbscan system. In each eye, 2 investigators performed 5 repeated measurements with each pachymetric device. Intraclass correlation coefficients (kappa) were calculated and mean CCT measurements were compared.

Main Outcome Measures: The CCT measurements obtained with ultrasound pachymetry, the Orbscan system (Orbtek Inc., Salt Lake City, UT), and PCI.

Results: Mean CCT values measured with ultrasound pachymetry were significantly thicker than those measured with PCI (21.5 microm; P<0.001) or the Orbscan system (19.8 microm; P<0.001). The correlation coefficients for the intraobserver variability were 0.999 for PCI measurements, 0.983 for ultrasound pachymetry measurements, and 0.988 for Orbscan system measurements. The correlation coefficients for the interobserver variability were 0.998 for PCI measurements, 0.980 for ultrasound pachymetry measurements, and 0.988 for Orbscan system measurements. There was a slightly better consistency between ultrasound pachymetry and PCI (kappa = 0.96) than between the Orbscan system and PCI (kappa = 0.92) and between ultrasound pachymetry and the Orbscan system (kappa = 0.89).

Conclusions: Partial coherence interferometry was the method with the least intraobserver or interobserver variability. Mean CCT as measured with ultrasound pachymetry was approximately 20 microm thicker than with the Orbscan system and PCI. However, corneal thickness measurements with ultrasound pachymetry and PCI were slightly more consistent than those of the Orbscan system and PCI. This slightly better consistency, however, may be important, especially in corneal refractive surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ophtha.2003.09.027DOI Listing

Publication Analysis

Top Keywords

ultrasound pachymetry
16
cct measurements
16
orbscan system
12
pachymetry orbscan
12
central corneal
8
corneal thickness
8
thickness measurements
8
partial coherence
8
coherence interferometry
8
pci ultrasound
8

Similar Publications

Purpose: To assess real-time intraoperative pachymetry changes occurring during corneal cross-linking (CXL) using microscope-integrated optical coherence tomography (Mi-OCT) and compare accuracy of various modalities of pachymetry assessment.

Methods: This was a cross-sectional observational comparative study including 45 consecutive patients with progressive keratoconus planned for CXL. Mi-OCT (RESCAN 700) was used to measure central corneal thickness (CCT) during four stages of CXL: before epithelial debridement, after debridement, after riboflavin instillation, and after ultraviolet irradiation.

View Article and Find Full Text PDF

Purpose: To evaluate the effect of eyelid laxity on the structure of the cornea.

Methods: This was a retrospective study performed in one institution. Patients with and without upper and lower eyelid laxity, which were determined using the snap-back test (SBT) and the upper eyelid distraction test (LDT), were evaluated for corneal structural changes with Scheimpflug imaging corneal tomography for the keratometric readings, and cor-neal spectral domain optical coherence tomography for the epithelial thickness (ET) measurements.

View Article and Find Full Text PDF

What intrinsic factors affect the central corneal thickness?

Ophthalmic Physiol Opt

January 2025

Physics Centre of Minho and Porto Universities (CF-UM-UP), School of Sciences, University of Minho, Braga, Portugal.

Article Synopsis
  • The cornea is essential for protecting the eye, and central corneal thickness (CCT) measures its thickness from the front to the back.
  • CCT is crucial for accurate intraocular pressure measurements, assessing corneal health, and planning refractive surgeries.
  • Understanding the various factors that affect CCT can improve diagnostic and treatment approaches in eye care.
View Article and Find Full Text PDF

Intrastromal Corneal Ring Segments and Keratoconus Progression: A Case Series Study.

Eye Contact Lens

February 2025

Cornea Unit (B.B.-P.), Hospital Severo Ochoa, Madrid, Spain; Department of Ophthalmology (P.A.-V., R.C.-S., D.D.-V.), Department of Immunology, Ophthalmology and ENT, Faculty of Medicine, Universidad Complutense de Madrid (UCM) and IdISSC, Hospital Clínico San Carlos, Madrid, Spain; Department of Ophthalmology (O.R.-Q.), Hospital Clínico San Carlos and IdISSC, Madrid, Spain; and Department of Surgical and Clinical, Diagnostic and Pediatric Sciences (V.T.), Section of Ophthalmology, University of Pavia, IRCCS, Policlinico San Matteo, Pavia, Italy.

Objective: To assess keratoconus (KC) progression following the implant of intrastromal corneal ring segments (ICRSs) in young patients.

Materials And Methods: Retrospective, longitudinal, observational, controlled nonrandomized case series study. Keratoconus patients aged 25 years or younger who underwent uneventful ICRS surgery and completed at least 3 years of follow-up were enrolled.

View Article and Find Full Text PDF

Purpose: To study the effect of corneal white-to-white diameter on Pentacam Scheimpflug cornea tomography and Corvis ST corneal visualization Scheimpflug technology parameters in myopic eyes.

Methods: In this retrospective cross-sectional study, 190 eyes of 190 Chinese myopic patients (spherical equivalent range: -1.25 D to -11.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!