Background And Purpose: To study the effect of breathing motion on gross tumor volume (GTV) coverage for lung tumors using dose-volume histograms and relevant dosimetric indices.

Patients And Methods: Treatment plans were chosen for 12 patients treated at our institution for lung carcinoma. GTV volumes of these patients ranged from 1.2 to 97.3 cm(3). A margin of 1-2 cm was used to generate the planning target volume (PTV). Additional margins of 0.6-1.0 cm were added to the PTV when designing treatment portals. For the purposes of TCP calculation, the prescription dose was assumed to be 70 Gy to remove the effects of prescription differences. Setup error was incorporated into the evaluation of treatment plans with a systematic component of sigma(RL) = 0.2 cm, sigma(AP) = 0.2 cm, and sigma(SI) = 0.3 cm and a random component of sigma(RL) = 0.3 cm, sigma(AP) = 0.3 cm, and sigma(SI) = 0.3 cm. Breathing motion was incorporated into these plans based on an independent analysis of fluoroscopic movies of the diaphragm for 7 patients. The systematic component of breathing motion (sigma(RL) = 0.3 cm, sigma(AP) = 0.2 cm, and sigma(SI) = 0.6 cm) was incorporated into the treatment plans on a slice by slice basis. The intrafractional component of breathing motion (sigma(RL) = 0.3 cm, sigma(AP) = 0.2 cm, and sigma(SI) = 0.6 cm) was incorporated by averaging the dose calculation over all displacements of the breathing cycle. Each patient was simulated 500 times to discern the range of possible outcomes. The simulations were repeated for a worst case scenario which used only breathing data with a large diaphragmatic excursion, both with and without intrafractional breathing motion.

Results: Dose to 95% of the GTV (D95), volume of the GTV receiving 95% of the prescription dose (V95) and TCP changed an average of -1.4+/-4.2, -1.0+/-3.3, and -1.4+/-3.8%, respectively, with the incorporation of normal breathing effects. In the worst case scenario (heavy breathers), D95 and V95 changed an average of -9.8+/-10.1 and -8.3+/-11.3%, respectively, and TCP changed by -8.1+/-9.1%. GTVs with volumes greater than 60 cm(3) showed stronger sensitivity to breathing especially if the shape was non-ellipsoidal. In the normal breathing case, the probability of a decrease in D95, V95, or TCP of a magnitude greater than 10% is less than 4%, and in the worse case scenario this probability is approximately 30-40% with intrafractional breathing motion included, and less than 10% with intrafractional breathing motion not included.

Conclusions: With the PTV margins routinely used at our center, the effects of normal breathing on coverage are small on the average, with a less than 4% chance of a 10% or greater decrease in D95, V95, or TCP. However, in patients with large respiration-induced motion, the effect can be significant and efforts to identify such patients are important.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2004.01.011DOI Listing

Publication Analysis

Top Keywords

breathing motion
24
sigmarl sigmaap
16
sigmaap sigmasi
16
breathing
13
treatment plans
12
case scenario
12
intrafractional breathing
12
v95 tcp
12
normal breathing
12
d95 v95
12

Similar Publications

Cine-magnetic resonance imaging (MRI) has been used to track respiratory-induced motion of the liver and tumor and assist in the accurate delineation of tumor volume. Recent developments in compressed sensitivity encoding (SENSE; CS) have accelerated temporal resolution while maintaining contrast resolution. This study aimed to develop and assess hepatobiliary phase (HBP) cine-MRI scans using CS.

View Article and Find Full Text PDF

Humidity sensors have been widely used to monitor humidity in daily life, agriculture fields, and so on. However, conventional sensors are not suitable for wearable devices because of their large dimensions and rigid substrates. Hence, we report a fast response, highly sensitive, and fully flexible humidity sensor on a PI substrate based on the composite material of reduced graphene oxide (rGO)/MoS, with a response time of 0.

View Article and Find Full Text PDF

Ultrasound (US) is a widely used technique for liver disease but has limitations in distinguishing tumors. This study evaluates the clinical efficacy of fluctuational imaging (FLI), a new US method that detects the fluttering sign in liver tumors. We conducted a prospective exploratory study with 120 participants diagnosed with liver tumors through histopathology or standard imaging.

View Article and Find Full Text PDF

Coupling the thermal acoustic modes of a bubble to an optomechanical sensor.

Microsyst Nanoeng

December 2024

ECE Department, University of Alberta, 9211-116 St. NW, Edmonton, T6G 1H9, AB, Canada.

Optomechanical sensors provide a platform for probing acoustic/vibrational properties at the micro-scale. Here, we used cavity optomechanical sensors to interrogate the acoustic environment of adjacent air bubbles in water. We report experimental observations of the volume acoustic modes of these bubbles, including both the fundamental Minnaert breathing mode and a family of higher-order modes extending into the megahertz frequency range.

View Article and Find Full Text PDF

Background: Cerebrospinal fluid (CSF) motion and pulsatility has been proposed to play a crucial role in clearing brain waste. Although its driving forces remain debated, increasing evidence suggests that large amplitude vasomotion drives such CSF fluctuations. Recently, a fast blood-oxygen-level-dependent (BOLD) fMRI sequence was used to measure the coupling between CSF fluctuations and low-frequency hemodynamic oscillations in the human cortex.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!