Purpose: To assess the performance of the Pascal dynamic contour tonometer (PDCT) (Swiss Microtechnology AG) by comparing it to Goldmann applanation tonometry (GAT) and noncontact air tonometry (NCT) before and after laser in situ keratomileusis (LASIK).
Setting: Vlemma Eye Institute, Athens, Greece.
Methods: Intraocular pressure was measured in 118 eyes before and 1 and 4 weeks after LASIK using GAT, NCT, and the PDCT, which allows direct, digital, transcorneal measurement of intraocular pressure.
Results: Preoperatively, central corneal thickness (CCT) correlated with GAT and NCT measurements but not with PDCT measurements. After LASIK, the mean reduction in CCT ranged from -3.0 to -171.0 microm (median 78 microm). The mean GAT measurement dropped by -4.9 mm Hg +/- 2.7 (SD) at 1 week and was -5.4 +/- 3.0 mm Hg at 4 weeks. Similar drops were observed in NCT measurements. The observed post-LASIK changes in GAT and NCT IOP measurements were not directly proportional to the change in CCT, refractive error, or mean keratometric readings. The preoperative and postoperative PDCT measurements did not differ significantly.
Conclusions: The structural corneal changes induced by LASIK appeared to influence GAT and NCT measurements but not PDCT measurements. Therefore, PDCT may be better suited for monitoring IOP in unoperated eyes and in eyes that have had LASIK.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcrs.2003.12.033 | DOI Listing |
Int Ophthalmol
June 2024
Department of Opthalmology, Kartal Dr. Lütfi Kırdar City Hospital, Kartal, Istanbul, Turkey.
Background: To evaluate the agreement between the Goldman applanation tonometer (GAT), Tono-Pen, and noncontact tonometer (NCT) in the measurement of intraocular pressure (IOP) in pseudophakic children.
Methods: The medical records of nonglaucomatous pseudophakic children between 2009 and 2019 were retrospectively analyzed. A total of 46 eyes of 23 patients operated for bilateral pediatric cataract were included in the study.
Introduction: Intraocular pressure (IOP) measurement in patients with thyroid-associated orbitopathy (TAO) can be difficult and misleading, particularly in patients with diplopia and eye deviation (esotropia or hypotropia). However, when measuring IOP, it is also necessary to pay sufficient attention to TAO patients without diplopia in primary gaze direction and without motility disorder that might not be readily apparent.
Purpose: The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using three different types of tonometers: the rebound tonometer (iCARE), the Goldmann applanation tonometer (GAT) and the non-contact airpuff tonometer (NCT) in patients with inactive TAO.
Int J Ophthalmol
October 2023
Department of Ophthalmology, Xijing Hospital, Fourth Military Medical University, Xi'an 710023, Shaanxi Province, China.
Aim: To access the agreement of intraocular pressure (IOP) values obtained from biomechanically corrected tonometer [Corvis ST (CST)], non-contact tonometer (NCT), and Goldmann applanation tonometer (GAT) in children with NCT measured-IOP (NCT-IOP) values of 22 mm Hg or more, and related factors.
Methods: A total of 51 eyes with NCT-IOP≥22 mm Hg in children aged 7 to 14y were examined and IOP was measured by CST, NCT, and GAT. Based on GAT measured IOP (GAT-IOP), ocular hypertension (OHT) group (≥22 mm Hg, 24 eyes) and the non-OHT group (<22 mm Hg, 27 eyes) were defined.
J Fr Ophtalmol
December 2023
Department of Ophthalmology, Erol Olcok Training and Research Hospital, Hitit University, Inönü avenue, N(o) 176, 19040, Çorum, Turkey. Electronic address:
Purpose: The goal of this paper was to compare the intraocular pressure (IOP) measurements obtained via iCare Pro rebound (IRT), non-contact tonometry (NCT), and Goldmann applanation (GAT) tonometry in healthy subjects.
Materials And Methods: One hundred and twenty-five healthy individuals were included in this study. The participants' IOP measurements were obtained via non-contact tonometry.
Indian J Ophthalmol
May 2023
Department of Cornea, Aravind Eye Hospital, Coimbatore, Tamil Nadu, India.
Purpose: To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT).
Methods: This was a prospective, cross-sectional, observational study to which patients above 18 years of age were enrolled. A total of 400 eyes of 200 non-glaucomatous patients underwent IOP recordings using the GAT, NCT, and RBT, and CCT was also noted.