The statistical characteristics of the intracavity laser spectrometer are investigated by analysis of cesium in aqueous solutions. The error distribution of measurements is normal. The relative standard deviation of the analysis in the middle of the dynamic range is equal to 9.7%. The intracavity spectral analysis was carried out at low contents of cesium, manganese and terbium in aqueous solutions. The limits of detection are: 5 microg l(-1) (Cs), 20 microg l(-1) (Mn) and 1000 microg l(-1) (Tb). Owing to the high spectral resolution of the spectrometer used the hyperfine structure of the thallium absorption line (535.05 nm) was recorded.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s0021663550317DOI Listing

Publication Analysis

Top Keywords

microg l-1
12
cesium manganese
8
manganese terbium
8
intracavity laser
8
aqueous solutions
8
determination cesium
4
terbium intracavity
4
laser spectroscopy
4
spectroscopy statistical
4
statistical characteristics
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!