Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Context: Current standards for laboratory accreditation from the College of American Pathologists state that when high-performance liquid chromatography (HPLC) is used as a screening test, all non-A, non-S abnormal hemoglobin (Hb) variants must be confirmed by an alternative method, including alkaline and acid electrophoresis.
Objective: To determine whether confirmation of Hb C and Hb O-Arab variants by an alternative method is required when using the Bio-Rad Variant II HPLC system.
Design: We reviewed 48 478 consecutive hemoglobin identification test results performed on the Bio-Rad Variant II HPLC system during the period November 15, 2000 to January 15, 2003.
Setting: Special Hematology Laboratory, Department of Pathology, Bellevue Hospital Center, New York, NY.
Main Outcome Measures: The chromatogram patterns and retention times (RTs) for specimens containing Hb C and Hb O-Arab were analyzed. We compared the results by the HPLC method with those by the confirmatory tests (alkaline and acid electrophoresis) for both variants.
Results: We identified 3668 cases of abnormal hemoglobin variants, including 660 cases of Hb C trait (17%), 5 cases of Hb O-Arab trait (0.1%), and 1 case of Hb SO-Arab (0.03%). A unique pattern of separation on the chromatogram for Hb O-Arab was revealed, presenting as 2 distinct peaks in 2 different manufacturer-defined RT windows, namely, D and C. The chromatogram for Hb C did not show the D window in any of the reviewed cases. The RT in the C window (C-RT) revealed a statistically significant difference for Hb C and Hb O-Arab (5.18 +/- 0.01 minutes and 4.91 +/- 0.01 minutes, respectively; P <.001).
Conclusion: According to our review, the identification of Hb C and Hb O-Arab is accurate using HPLC methodology, as performed by the Bio-Rad Variant II HPLC system. This method can be both confirmatory and diagnostic at the same time.