Background: Urinary excretion of the low molecular weight protein cystatin C is a marker of renal disorders and a good predictor of the severity of acute tubular necrosis. We evaluated the measurement of urinary cystatin C and determined its reference range.

Methods: Measurement of urinary cystatin C by particle-enhanced nephelometric immunoassay (PENIA) in 102 patients with various renal disorders and 133 healthy controls. We assessed the influence of pH and temperature, interferences on urinary cystatin C measurement, as well as cystatin C adsorption to plastic.

Results: The upper reference value for urinary cystatin C was 0.28 mg/L, independent of age and gender. Accuracy and linearity (r(2)=0.996) were excellent. Intra- and inter-assay precision were < or =4.8% and < or =5.2%, respectively. Albumin (< or =160 g/L), bilirubin (< or =500 micromol/L) and haemoglobin (< or =210 micromol/L) did not show interferences. Urinary cystatin C was stable, at urine pH> or =5, at -20 degrees C and 4 degrees C for 7 days, and at 20 degrees C for 48 h. Freezing and thawing did not influence urinary cystatin C concentration. There was no adsorption of cystatin C to plastic.

Conclusion: Urinary cystatin C measurement by PENIA is precise. High stability and no interference add to the practicability of urinary cystatin C as a routine biochemical test.

Download full-text PDF

Source
http://dx.doi.org/10.1258/000456304322879980DOI Listing

Publication Analysis

Top Keywords

urinary cystatin
36
measurement urinary
12
cystatin
12
urinary
9
cystatin particle-enhanced
8
particle-enhanced nephelometric
8
nephelometric immunoassay
8
renal disorders
8
interferences urinary
8
cystatin measurement
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!