Transvenous unipolar active can defibrillation systems have proven to be effective in treating ventricular tachyarrhythmias. However, a further reduction of ventricular defibrillation thresholds (V-DFT) would increase the longevity, reduce the size of pulse generators, and help to avoid additional leads in patients with inacceptable high V-DFTs. In a finite difference computer model, the extension of the right ventricular (RV) defibrillation coil into the low right atrium led to a 40% reduction of unipolar V-DFT. To evaluate this finding, we conducted a prospective, randomized study in 11 patients receiving an ICD. Extension of the RV electrode was simulated by adding a second coil placed in the low right atrium with the same polarity. Using a binary search protocol, V-DFT was determined with and without the additional electrode in each patient. Total shock impedance was significantly lower in the two coil (low RA) configuration, compared to the single coil (RV) configuration. Corresponding values were 49.9 +/- 6.7 Ohm and 61.1 +/- 9.3 Ohm, respectively (P < 0.01, paired t-test). However, there was no reduction, but even a nonsignificant increase in V-DFTs. Mean V-DFT in the RV configuration was 12.0 +/- 5.6 J and 16.3 +/- 7.8 J in the low RA configuration (P = 0.09, paired t-test). Despite a reduction in total impedance, the addition of a defibrillation coil in the low right atrium does not reduce ventricular defibrillation thresholds.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2004.00439.xDOI Listing

Publication Analysis

Top Keywords

low atrium
16
ventricular defibrillation
16
coil low
16
defibrillation thresholds
12
addition defibrillation
8
reduce ventricular
8
defibrillation coil
8
low configuration
8
+/- ohm
8
paired t-test
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!