Objective: To qualitatively validate radioimmunoassays for 8-isoprostane and prostaglandin (PG) E(2) in exhaled breath condensate.

Subjects: Twenty-two subjects with different lung diseases attended the outpatient clinic on one occasion for exhaled breath condensate collection.

Methods: Samples were pooled together and purified by reverse phase high performance liquid chromatography (RP-HPLC). The eluted fractions were assayed for 8-isoprostane-like immunoreactivity and PGE(2)-like immunoreactivity by radioimmunoassays. In addition, simultaneous measurements of exhaled breath condensate unextracted samples with two anti-8-isoprostane and anti-PGE(2) sera with different cross-reactivity were performed.

Results: A single peak of 8-isoprostane-like immunoreactivity and PGE(2)-like immunoreactivity co-eluting with 8-isoprostane (retention time: 13 min) and PGE(2) (retention time: 21 min) standards, respectively, was identified by radioimmunoassays. Testing with two different antisera showed similar results for both 8-isoprostane-like immunoreactivity (limits of agreement = 4.5 pg/ml and - 4.1 pg/ml, n = 12) and PGE(2)-like immunoreactivity (limits of agreement = 6.1 pg/ ml and - 6.1 pg/ml, n = 12).

Conclusion: This study provides evidence for the specificity of the radioimmunoassays for 8-isoprostane and PGE(2) in exhaled breath condensate. This is critical for proposing these markers as a non-invasive way for monitoring airway inflammation.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00011-003-1212-6DOI Listing

Publication Analysis

Top Keywords

exhaled breath
20
breath condensate
16
8-isoprostane-like immunoreactivity
12
pge2-like immunoreactivity
12
8-isoprostane prostaglandin
8
measurements exhaled
8
radioimmunoassays 8-isoprostane
8
immunoreactivity pge2-like
8
retention time
8
time min
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!