The ability of the conductance catheter method to measure left ventricular segmental and total volume was evaluated by comparison with the Cine-CT technique. In the seven dogs studied, 19 conductance catheter and simultaneous Cine-CT runs were obtained. High correlation coefficients were found for total volume and segmental volumes, except in the basal segment. However, in most cases there was a significant variability in slope and intercept between animals. Both methods are promising tools for estimating dynamic segmental left ventricular volume, each having specific advantages such as a continuous signal (conductance catheter) or anatomic detail (Cine-CT). However, the results also show the need for further improvement of both methods.

Download full-text PDF

Source
http://dx.doi.org/10.1093/eurheartj/13.suppl_e.15DOI Listing

Publication Analysis

Top Keywords

conductance catheter
16
left ventricular
12
ventricular segmental
8
total volume
8
segmental
4
volume
4
segmental volume
4
conductance
4
volume conductance
4
catheter
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!