[A study of 11 ventilators for anesthesia: laboratory testing].

Ann Fr Anesth Reanim

Service d'Anesthésie-Réanimation, Hôpital Henri-Mondor, Créteil.

Published: February 1993

Eleven anaesthesia ventilators were instrumentally tested under various conditions. They included: Excel and Modulus II Plus (Ohmeda); 710 and Servo anaesthesia circle 985 (Siemens); Jollytronic (Soxil) and Elsa (Engström); SA2 and Cicero (Dräger); ABT 4,300 (Kontron); Monnal A and the prototype Alys (Taema). The test circuit comprised a two compartment model lung, a pneumotachograph, a pressure gauge in the "airway". The volume was calculated as the integral of flow rate. Each machine was calibrated by the firms' technicians. Before each test, the pneumotachograph was calibrated using a 11 air syringe and the pressure gauge with a 5 cm water column. Each machine ventilated the model lung for 30 min before starting the tests. There were five tests: 1) reliability of the machine's spirometer, 2) reliability of the ventilation rate, 3) reliability of pressure measurements, 4) effect of increasing fresh gas flow on spirometry, 5) effect of increasing downstream resistances. In usual simulated ventilatory conditions, all the machines accurately delivered the setted ventilation and spirometric measurements were with minimal error only. Several ventilators (SA2, Excel, 710, Elsa, ABT 4,300) did not succeed in maintaining their performances when compliance was strongly decreased or resistance of the test lung notably increased. Resistance in the circuit during simulated spontaneous ventilation was < 3.6 cmH2O.l-1.s-1. Increasing fresh gas flow raised the minute volume delivered in six ventilators. It is concluded that, during extreme ventilatory conditions, the inspired volume must be adjusted so as to maintain the inspired tidal volume. However, ventilators which increase inspiratory time in response to an increased mechanical load cannot be adjusted by this way.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0750-7658(05)80755-5DOI Listing

Publication Analysis

Top Keywords

abt 4300
8
model lung
8
pressure gauge
8
increasing fresh
8
fresh gas
8
gas flow
8
ventilatory conditions
8
study ventilators
4
ventilators anesthesia
4
anesthesia laboratory
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!