Introduction: Characterization of the performance of an hyperthermia applicator by phantom experiments is an essential aspect of quality assurance in hyperthermia. The objective of this study was to quantitatively characterize the energy distribution of the Sigma-60 applicator of the BSD2000 phased array system operated within the normal frequency range of 70-120 MHz. Additionally, the accuracy of the flexible Schottky diode sheet to measure E-field distributions was assessed.
Material And Methods: The flexible Schottky diode sheet (SDS) consists of 64 diodes mounted on a flexible 125 microm thick polyester foil. The diodes are connected through high resistive wires to the electronic readout system. With the SDS E-field distributions were measured with a resolution of 2.5 x 2.5 cm in a cylindrical phantom, diameter of 26 cm and filled with saline water (2 g/l). The phantom was positioned symmetrically in the Sigma-60 applicator. RF-power was applied to the 4-channel applicator with increasing steps from 25W to a total output of 400 W.
Results: The complete system to measure the E-field distribution worked fine and reliably within the Sigma-60 applicator. The E-field distributions measured showed that the longitudinal length of the E-field distribution is more or less constant, e.g. 21-19 cm, over the frequency range of 70-120 MHz, respectively. As expected, the radial E-field distributions show a better focusing towards the centre of the phantom for higher frequencies, e.g. from 15.3-8.7 cm diameter for 70-120 MHz, respectively. The focusing target could be moved accurately from the left to the right side of the phantom. Further it was found that the sensitivity variation of nine diodes located at the centre of the phantom was very small, e.g. < 3% over the whole frequency range.
Conclusion: The SAR distributions of the Sigma-60 applicator are in good agreement with theoretically expected values. The flexible Schottky diode sheet proves to be an excellent tool to make accurate, quantitative measurements of E-field distributions at low (25 W) and medium (400 W) power levels. An important feature of the SDS is that it enables one to significantly improve quantitative quality assurance procedures and to start quantitative comparisons of the performance of the different deep hyperthermia systems used by the various hyperthermia groups.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/0265673031000140813 | DOI Listing |
Int J Hyperthermia
December 2024
Department of Electrical Engineering, Chalmers University of Technology, Gothenburg, Sweden.
Radiat Oncol
July 2015
Department of Radiation Oncology, Charité Universitätsmedizin Berlin, Augustenburger Platz 1, 13353, Berlin, Germany.
Background: Peritoneal carcinomatosis occurs in different cancer subtypes and is associated with a dismal prognosis. Some doubts remain whether the whole abdomen can be treated by regional hyperthermia, therefore we analyzed feasibility conducting a pilot study.
Methods: A simulation of the abdominopelvic heat distribution in 11 patients with peritoneal carcinomatosis was done using the HyperPlan software and the SIGMA-60 and SIGMA-Eye applicators.
Int J Hyperthermia
May 2014
Medical Physics Graduate Program, Duke University, Durham , North Carolina .
Purpose: The aim of this paper is to report thermal dosimetry characteristics of external deep regional pelvic hyperthermia combined with intravesical mitomycin C (MMC) for treating bladder cancer following transurethral resection of bladder tumour, and to use thermal data to evaluate reliability of delivering the prescribed hyperthermia dose to bladder tissue.
Materials And Methods: A total of 14 patients were treated with MMC and deep regional hyperthermia (BSD-2000, Sigma Ellipse or Sigma 60). The hyperthermia objective was 42° ± 2 °C to bladder tissue for ≥40 min per treatment.
Strahlenther Onkol
January 2013
Department of Radiation Oncology, Erasmus MC - Daniel den Hoed, Groene Hilledijk 301, 3008 AE, Rotterdam, The Netherlands.
Background And Purpose: To investigate the clinical benefit of replacing the BSD-2000 Sigma-60 with the Sigma-Eye applicator, taking into account effects of uncertainties in tissue and water bolus parameters.
Patients And Methods: For 20 patients, specific absorption rate (SAR) and temperature distributions were calculated and optimized, based on computed tomography (CT) scans in treatment position. The impact of uncertainties on predicted distributions was studied using a Monte Carlo uncertainty assessment.
Int J Hyperthermia
August 2011
Department of Medical Physics, Faculty of Medicine, Shahrekord University of Medical Sciences, Shahrekord, Iran.
Purpose: To investigate whether one can replace the Sigma-60 with the Sigma-Eye applicator (or vice versa) during a deep hyperthermia treatment series without a loss in quality of the treatment.
Patients And Methods: Hyperthermia data of all 48 patients with locally advanced cervical cancer who were treated with both applicators were analysed. In this study no use was made of the longitudinal SAR steering option of the Sigma-Eye.