Unlabelled: Parkinson's disease patients with long-term L-dopa syndrome may benefit from an implanted cerebral stimulation device. When advanced life support demands cardioversion or defibrillation in these patients, undesired effects of monophasic electroshocks might occur in brain tissue adjacent to the stimulation electrodes (e.g., thermal injury), but also in the stimulation device itself. Thus, in this animal study (n = 6 pigs), we investigated the effects of repeated defibrillation (2 x 200 J [n = 1] and 2 x 360 J [n = 5]) at the implantation site of cerebral stimulation electrodes and on stimulation device function. Repeated external cardiac defibrillation did not cause acute histopathologic changes typical of thermal injury to brain tissue adjacent to the cerebral stimulation electrodes. Functionality of the stimulator device after defibrillation, however, ranged from normal to total loss of function. Therefore, when defibrillation is performed, the greatest possible distance between the defibrillation site and the stimulator device implantation site should be considered. Subsequent testing of the stimulator device's function is mandatory.

Implications: Repeated cardiac defibrillation did not cause histopathologic changes typical of thermal injury at the implantation site of cerebral stimulation electrodes. The function of the stimulator device after defibrillation, however, ranged from normal to total loss of function.

Download full-text PDF

Source
http://dx.doi.org/10.1213/01.ANE.0000096192.33388.48DOI Listing

Publication Analysis

Top Keywords

cerebral stimulation
20
stimulation electrodes
20
cardiac defibrillation
12
changes typical
12
typical thermal
12
stimulation device
12
thermal injury
12
implantation site
12
stimulator device
12
defibrillation
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!