Background: The use of nonmydriatic cameras, which offer ease of screening and 45 degrees immediate imaging of the fundus, is gaining increasing acceptance for screening programs tailored to diverse conditions. We performed a study to evaluate the effectiveness and safety of screening for diabetic retinopathy with two nonmydriatic camera images compared with the seven standard stereoscopic 30 degrees fields (7SF). We also wished to determine whether safe screening guidelines could be established to identify patients needing referral to an ophthalmologist.

Methods: In this prospective masked cross-sectional study, we evaluated agreement in the assessment of the severity of diabetic retinopathy by means of two 45 degrees images centred on the optic disc and on the macula obtained with the Topcon CRW6 nonmydriatic camera and by means of 7SF photography and ophthalmologic slit-lamp biomicroscopy, both performed with pupil dilation. Between November 2000 and June 2001, 98 adult patients known to have type 1 or 2 diabetes mellitus who presented for the first time to the diabetic retinopathy clinic of a tertiary care centre in Montreal were enrolled consecutively. Thus, patient recruitment was weighted toward more severe retinopathy to ensure sufficient representation of less frequent but more severe levels. Each patient underwent nonmydriatic fundus photography of both eyes, followed by a complete ophthalmologic examination with pupil dilation by a single retina specialist and 7SF photography of both eyes with pupil dilation. The level of retinopathy was graded independently in each eye from the 7SF photographs according to the Early Treatment Diabetic Retinopathy Study (ETDRS) scale by two graders; an independent retina specialist adjudicated the rare instances of interreader disagreement in a masked fashion. Two months later, two graders independently graded the nonmydriatic images in a blinded fashion according to the ETDRS scale; a third observer adjudicated the rare instances of interreader disagreement. We measured concordance between grading results with the various screening techniques using the weighted and unweighted kappa statistic. We used sensitivity and specificity indices to determine safe screening guidelines to identify patients needing referral to an ophthalmologist.

Results: There was substantial agreement in the grading of retinopathy with nonmydriatic camera imaging and with 7SF photography, both for all eyes (kappa = 0.626 [standard deviation (SD) 0.045]) and for the eye with more severe disease (kappa = 0.654 [SD 0.063]). With nonmydriatic camera imaging, screening thresholds for patient referral to an ophthalmologist of very mild retinopathy (ETDRS grade 20), mild retinopathy (ETDRS grade 35) and moderate retinopathy (EDTRS grade 43) had sensitivity values of 97.9%, 97.1% and 53.3% respectively and specificity values of 81.3%, 95.5% and 96.9% respectively. Screening thresholds of very mild or mild retinopathy both correctly identified 100% of eyes with severe nonproliferative or proliferative retinopathy. With a screening threshold of mild retinopathy, screening with the nonmydriatic camera would lead to referral to an ophthalmologist of 37.8% of patients because of detected disease and of an additional 17.3% because of insufficient image quality in at least one eye, for a total of 55.1%. The overall sensitivity and specificity of a two-field nonmydriatic screening strategy with a threshold of mild retinopathy for referral of patients with insufficient image quality in at least one eye are 97.7% and 84.0% respectively.

Interpretation: Our results suggest that two-field nonmydriatic camera imaging is a safe screening strategy that may identify the patients with diabetes most in need of ophthalmologic care.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0008-4182(03)80109-6DOI Listing

Publication Analysis

Top Keywords

nonmydriatic camera
24
diabetic retinopathy
20
mild retinopathy
20
retinopathy
15
screening
13
retinopathy nonmydriatic
12
safe screening
12
identify patients
12
7sf photography
12
pupil dilation
12

Similar Publications

Cardiovascular diseases are a leading cause of death, and psychosocial stress is considered a contributing factor to these issues. With the rising number of heart surgeries, proper rehabilitation post-surgery is essential. Previous studies have demonstrated the positive impact of yoga and transcendental meditation on the cardiovascular system.

View Article and Find Full Text PDF

Increased intracranial pressure (ICP) is encountered in numerous traumatic and non-traumatic medical situations, and it requires immediate recognition and attention. Clinically, ICP typically presents with a headache that is most severe in the morning, aggravated by Valsalva-like maneuvers, and associated with nausea or vomiting. Papilledema is a well-recognized sign of increased ICP; however, emergency physicians often find it difficult to visualize the optic disc using ophthalmoscopy or to accurately interpret digital fundus photographs when using a non-mydriatic retinal camera.

View Article and Find Full Text PDF

Widefield fundus photography is critical for the detection, documentation, and management of pediatric eye diseases. Existing clinical pediatric fundus cameras offer a limited field of view (FOV) and suboptimal image contrast, hindering comprehensive peripheral retina examination. Additionally, the high cost and lack of portability of commercial devices restrict their use in resource-limited settings.

View Article and Find Full Text PDF

Purpose: Ophthalmology is frequently consulted to "rule-out papilledema." We assessed the efficacy of a portable, non-mydriatic fundus camera in detecting optic disc edema in the inpatient and emergency room settings during "rule-out papilledema" consultations, comparing to the gold standard dilated fundus examination (DFE).

Methods: This prospective, blinded, cohort study included 124 non-mydriatic fundus photographs (62 patients) that were obtained over the 11-month enrollment period.

View Article and Find Full Text PDF
Article Synopsis
  • - The study aimed to assess the safety and effectiveness of the Eidon white light 60-degree field Scanning Confocal Ophthalmoscope (SCO) camera for diabetic eye screening using staged mydriasis, involving a large trial with 1015 patients.
  • - Results showed high sensitivity (97.5%) and specificity (82.3%) for detecting any retinopathy, indicating the camera's reliability, although some images remained unassessable even after dilation.
  • - The researchers concluded that while the Eidon camera demonstrates promising performance, further studies are needed for diverse populations and to evaluate its cost-effectiveness.
View Article and Find Full Text PDF