[Evaluation of the heterogeneous immunoassay (ACMIA) for the measurement of blood cyclosporin on the Behring dimension RXL clinical chemistry analyzer].

Ann Biol Clin (Paris)

Laboratoire de pharmacologie et toxicologie, CHU Henri Mondor, 51, avenue du Maréchal de Lattre de Tassigny, 94010 Créteil.

Published: March 2004

We propose an evaluation of a new heterogeneous immunoassay of cyclosporin on RXL HM Dimension (Dade Behring) for therapeutic cyclosporin monitoring in whole-blood patients transplant. The pretreatment step is performed automatically into the apparatus while it is a manual step with EMIT. Linearity, intra- and inter-day precision, limit of quantification, precision and accuracy of dilution steps and stability into the equipment were studied. We realized the comparison between ACMIA and EMIT methods on whole-blood patients transplant recipients. Heterogeneous immunoassay showed a good linearity between 0 and 500 ng/mL, intra- and inter-day precision with coefficient of variation inferior to 7.2%. We observed reproducible and accurate dilutions of high concentrations (500 to 2,000 ng/mL). The correlation with EMIT technique was correct for different type of transplant (n=55).

Download full-text PDF

Source

Publication Analysis

Top Keywords

heterogeneous immunoassay
12
whole-blood patients
8
patients transplant
8
intra- inter-day
8
inter-day precision
8
[evaluation heterogeneous
4
immunoassay acmia
4
acmia measurement
4
measurement blood
4
blood cyclosporin
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!