Background: This study examined the effect of measuring blood pressure below subjects' rolled-up sleeves, over the sleeve, or on the bare arm. This is an important day-to-day issue for the busy GP.
Methods: The sample consisted of 201 subjects in family practice clinics and residents of a senior citizens' home. A digital device was used in all cases. Each participant underwent three blood pressure measurements in each of the following conditions in random order: cuff on bare arm; cuff over the sleeve; and cuff below the rolled-up sleeve. Differences between measurements were plotted against the mean blood pressure. Confounding factors controlled for were age, sex, clothing pressure and skin-fold thickness.
Results: Differences in mean blood pressure readings between the clothed and bare arm were 0.5 mmHg (SD 7.5) for systolic pressure and 1 mmHg (SD 5) for diastolic pressure; neither difference was significant. However, in hypertensive subjects (>140 mmHg systolic), although the mean difference remained small (systolic pressure, 2 mmHg, SD 10), the range of difference for individual subjects was -32 mmHg to +22 mmHg.
Conclusion: The degree of clothing under the sphygmomanometer cuff does not have a clinically important effect on the blood pressure measurement. In patients known or found to be hypertensive, measurement on the bare arm is recommended.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/fampra/cmg618 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!