A simple method of solventless extraction of volatile organic compounds (BTEX) from aqueous samples was developed and validated. A new arrangement of the full volume inside needle capillary adsorption trap (INCAT) device with Porapak Q as a sorbent material and wet alumina as a source of desorptive water vapour flow in a closed analytical system is presented. The analytical characteristics of developed device and of compared purge-and-trap (PTI) device for BTEX compounds are similar; the limits of detection as well as quantification are lower than 1 microg l(-1).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jchromb.2003.09.009 | DOI Listing |