Background: Pacing leads with a small electrode surface for high-impedance stimulation have been shown to prolong pacemaker longevity, but no sufficient data is available on the safety and feasibility of a defibrillation lead with this novel design.
Methods: We evaluated the clinical performance of a tined, steroid-eluting defibrillation lead with a small electrode surface area (model 6944) in a prospective multicenter study. A total of 542 patients with conventional indications for an implantable cardioverter defibrillator were randomized 1:1 to receive either the model 6944 or a tined, steroid-eluting defibrillation lead with a conventional sized electrode surface area (model 6942). Device performance and electrical parameters were evaluated at implant and 1, 3, 6, and 12 months thereafter (mean follow-up 11.3 +/- 5.6 months).
Results: Baseline characteristics, lead implant success rates, and defibrillation thresholds did not differ significantly between the 2 groups. While pacing thresholds did not differ significantly during follow-up, pacing impedance was approximately twice as high in the model 6944 as in the model 6942 lead (P <.0001). Mean R-wave amplitudes were smaller in patients with a 6944 (9.1 +/- 3.1 mV vs 9.8 +/- 3.6 mV for model 6942, P <.05), but remained stable within both groups throughout the observation period. The total number of ventricular lead-related adverse events and patient survival did not differ significantly between the 2 groups.
Conclusions: The use of a defibrillation lead with a small electrode surface for high-efficiency pacing is safe and feasible and increases pacing impedance without significantly compromising clinical performance.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/S0002-8703(03)00411-3 | DOI Listing |
Sci Rep
December 2024
Department of Medical and Surgical Sciences, Institute of Cardiology, University of Bologna, Policlinico S.Orsola-Malpighi, via Massarenti 9, Bologna, 40138, Italy.
Cardiac implantable electronic devices infections (CIEDI) are associated with poor survival despite the improvement in transvenous lead extraction (TLE). Aetiology and systemic involvement are driving factors of clinical outcomes. The aim of this study was to explore their contribute on overall mortality.
View Article and Find Full Text PDFWorld J Cardiol
December 2024
Department of Cardiovascular Medicine, Affiliated Hospital of Inner Mongolia University for Nationalities, Tongliao 028000, Inner Mongolia Autonomous Region, China.
Background: Ventricular arrhythmia is a common type of arrhythmia observed in clinical practice. It is primarily characterized by premature ventricular contractions, ventricular tachycardia, and ventricular fibrillation. Abnormal formation or transmission of cardiac electrical impulses in patients affects cardiac ejection function.
View Article and Find Full Text PDFJ Cardiovasc Dev Dis
December 2024
Cardiac Electrophysiology Division, Cardiology Center, Department of Internal Medicine, University of Szeged, 6725 Szeged, Hungary.
Background: An atrioventricular defibrillator system with a floating atrial dipole (VDD ICD) can provide atrial sensing by a single lead. Our aim was to compare the arrhythmia detection efficacy of VDD ICDs with conventional single- (VVI) and dual-chamber (DDD) defibrillators.
Methods: Data from consecutive patients undergoing ICD implantation were retrospectively analyzed.
ESC Heart Fail
December 2024
Boston Scientific Corporation, St. Paul, Minnesota, USA.
Pan Afr Med J
December 2024
Service d'Epidémiologie et Médecine Communautaire, Centre Hospitalier Universitaire Hedi Chaker de Sfax, Sfax, Tunisie.
The benefits of permanent cardiac pacing have been widely demonstrated. However, the literature on complications remains inconsistent. We lack precise information about the frequency of complications and their predictive factors in our center.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!