A bellows-based breath-hold monitoring and feedback system was developed and evaluated for use in intermittent-mode computed tomographic (CT) fluoroscopy-guided biopsy procedures in the lung or upper abdomen. The bellows system is described, and its feasibility is demonstrated in studies with a respiratory phantom and human volunteers. Results are reported for seven patients who underwent bellows-assisted biopsy. Breath-hold monitoring and feedback with the bellows system allow the patient to perform reliable breath holding at a preselected level. This optimizes intermittent-mode CT fluoroscopy-guided biopsies by allowing consistent visualization of the target lesion throughout the procedure.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2293021496DOI Listing

Publication Analysis

Top Keywords

breath-hold monitoring
12
monitoring feedback
12
intermittent-mode fluoroscopy-guided
8
fluoroscopy-guided biopsy
8
lung upper
8
upper abdomen
8
feedback system
8
bellows system
8
biopsy lung
4
abdomen breath-hold
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?