[Particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein].

Hunan Yi Ke Da Xue Xue Bao

Department of Clinical Laboratory, Second Xiangya Hospital, Central South University, Changsha 410011, China.

Published: August 2003

Objective: To evaluate the particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein by the automated chemistry analyzer (Hitachi 7170A).

Methods: The detection limit of quantification, measurement range, precision, interference and reference interval were assessed.

Results: The measurement range of the method was from 0.125 to 12.5 mg/L. The detection limitation was 0.115 mg/L. The internal and external accuracy were 0.89% and 0.90% respectively. The reference range of the assay was from 0.20 mg/L to 2.60 mg/L.

Conclusion: The particle enhanced immunoturbidimetic assay is sensitive and stable, and can be used in clinical diagnosis.

Download full-text PDF

Source

Publication Analysis

Top Keywords

enhanced immunoturbidimetic
12
immunoturbidimetic assay
12
assay high
8
high sensitive
8
sensitive c-reactive
8
particle enhanced
8
measurement range
8
[particle enhanced
4
assay
4
c-reactive protein]
4

Similar Publications

[Particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein].

Hunan Yi Ke Da Xue Xue Bao

August 2003

Department of Clinical Laboratory, Second Xiangya Hospital, Central South University, Changsha 410011, China.

Objective: To evaluate the particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein by the automated chemistry analyzer (Hitachi 7170A).

Methods: The detection limit of quantification, measurement range, precision, interference and reference interval were assessed.

Results: The measurement range of the method was from 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!