Long-term results of a cementless knee prosthesis with a metal-backed patellar component: clinical and radiological follow-up with histology from retrieved components.

J Long Term Eff Med Implants

Department of Biomaterials/Handicap Research, Sahlgrenska University Hospital, Institute for Surgical Sciences, University of Gothenburg, Gothenburg, Sweden.

Published: January 2004

Aseptic loosening of cemented knee arthroplasties has encouraged development of uncemented fixation. The Miller-Galante I (MG I) prosthesis was designed to achieve permanent stability through ingrowth into a titanium fiber mesh. Thirty-five knees in 30 patients with MG I knee replacements have been followed clinically and radiologically with a mean follow-up of 12 years. Twenty knees were revised with a mean follow-up of 5 years. Patellofemoral problems, especially avulsion of the polyethylene from the metal-backed patella and in some cases severe metallosis, have been the main reason for revision. Metal-backed patellar component should be avoided. In addition, instability has been a problem. However, the fixation of the components has been excellent, with a high degree of osseous ingrowth displayed at histological analysis of retrieved components. The clinical and radiological results after revision are in most cases good, despite the femoral component having been left in situ.

Download full-text PDF

Source
http://dx.doi.org/10.1615/jlongtermeffmedimplants.v13.i4.70DOI Listing

Publication Analysis

Top Keywords

metal-backed patellar
8
patellar component
8
clinical radiological
8
retrieved components
8
follow-up years
8
long-term cementless
4
cementless knee
4
knee prosthesis
4
prosthesis metal-backed
4
component clinical
4

Similar Publications

Early Shear Failure of a 3-Peg Modified Dome Patellar Implant.

Arthroplast Today

August 2024

Department of Orthopaedic Surgery, UCLA Health, Santa Monica, CA, USA.

Patellar failure in total knee arthroplasty is a major source of complications postoperatively. Previous patellar failure reports commonly cited dissociations of modular and metal-backed patellar implants. However, mechanical breakage of monoblock all-polyethylene patellar implants is very rare.

View Article and Find Full Text PDF

Metallosis is a known yet rare late complication of unicompartmental and total knee arthroplasty (TKA), usually secondary to either metal-backed patellar component failure, mobile-bearing polyethylene dislocation, or catastrophic polyethylene failure and wear through. The majority of literature surrounding metallosis has been published in relation to total hip arthroplasty (THA) metal on metal bearing wear or mechanically assisted crevice corrosion.This case report describes the development of metallosis in a 77-year-old male patient with advanced (Kellgren-Lawrence Grade 4) osteoarthritis with associated valgus deformity, who underwent index TKA with a semiconstrained revision knee system due to intraoperative medial collateral ligament laxity.

View Article and Find Full Text PDF

Aims: The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

View Article and Find Full Text PDF

Background: Cementless total knee arthroplasty (TKA) has become increasingly popular. Some surgeons are concerned about pain, implant stability, and metal-backed patellar component survivorship. This study investigated the outcomes of cementless compared with cemented TKA in bilateral cases.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!