Objectives: To compare arterial bursting pressure after vessel closure using a vessel-sealing device (LigaSure Atlas Laparoscopic Sealer/Divider Instrument; Valleylab, Boulder, CO), a ligate-and-divide stapling device (LDS), and 2-0 polydioxanone suture. To evaluate the LigaSure Atlas as a method for ligation of the mesenteric vasculature during small intestinal resection in normal horses.
Study Design: Experimental study.
Animals: Part A: jejunal segments from 19 horses. Part B: 6 horses, aged 1 to 18 years, weighing 330 to 509 kg.
Methods: Part A: Jejunal segments with mesenteric vessels were collected from 19 horses. After closure by 1 of 3 methods (LigaSure Atlas, LDS, 2-0 polydioxanone) arteries were cannulated, and bursting pressure was measured by incrementally increasing intraluminal pressure until failure. Part B: Six horses had jejunal resection and anastomosis using a vessel-sealing device (LigaSure Atlas) to provide hemostasis of the mesenteric vasculature. Horses were monitored clinically for 4 weeks.
Results: Part A: Mean +/- SEM bursting pressure after 2-0 polydioxanone ligation (1,014.50 +/- 279.05 mm Hg) was significantly greater than mean bursting pressure after LigaSure (554.25 +/- 228.79 mm Hg), which was significantly greater than the mean bursting pressure after LDS (373.25 +/- 183.69). Part B: No major operative or postoperative hemorrhage occurred after application of the LigaSure Atlas for sealing and transecting mesenteric vasculature during small intestinal resection in normal horses.
Conclusions: The LigaSure Atlas appears to be a safe method for hemostasis of the mesenteric vasculature during small intestinal resection in normal horses.
Clinical Relevance: Benefits of the LigaSure Atlas vessel-sealing device include reduced time required to provide hemostasis, acceptable arterial bursting pressure, no remaining foreign material, and no risk for ligature slippage. In our experience, use of the LigaSure Atlas during small intestinal resection and anastomosis in horses is safe for ligation of vessels less than or equal to 7-mm diameter.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1532-950x.2003.00574.x | DOI Listing |
Medicina (Kaunas)
May 2023
Department of Pediatric Surgery, Pediatric Trauma and Pediatric Urology, University Medical Centre Magdeburg, 39120 Magdeburg, Germany.
: A topic of greatinterest in the surgical field comprises cost and time reduction operative techniques with high efficiency rates. Thus, the aim of this paper is to evaluate whether a transection of the appendix using only a laparoscopic LigaSure™ device is feasible and, if so, which size of the laparoscopic device is optimal. : Appendectomy specimens were sealed and cut using LigaSure V (5 mm) and LigaSure Atlas (10 mm) devices ex vivo.
View Article and Find Full Text PDFAnimals (Basel)
September 2022
Dipartimento Dell'Emergenze e Trapianti di Organo, sez. Cliniche Veterinarie e P.A., Università Degli Studi di Bari, s.p. per Casamassima km.3, Valenzano, 70010 Bari, Italy.
To evaluate two vessel-sealing devices with different jaw lengths for dissecting ovaries in adult obese African lionesses undergoing laparoscopic ovariectomy. Twelve lionesses ( = 12) were recruited. The surgical procedures were performed through three portals, with a retractor platform positioned at the umbilical port and cannulas placed 3-4 cm cranial and caudal to the device at the level of the midline.
View Article and Find Full Text PDFVet Sci
August 2022
Department of Emergencies and Organ Transplantation, Section of Veterinary Clinics and Animal Production, University of Bari, Valenzano, 70010 Bari, Italy.
This study compared burst pressure (BP), number of activations, and histological assessment of ex vivo swine small intestine loops transected by stapler, a single fulcrum radiofrequency vessel sealing (RFVS) device, and the newly-developed jaws RFVS. Fifty (n = 50) 20 cm long jejunal loops were randomly assigned to be transected with RFVS devices and linear stapler (Caiman5, Caiman Maryland, Caiman12, Ligasure Atlas, and Stapler group as control respectively). Caiman5, Caiman12 and stapler required only one activation to complete the sealing.
View Article and Find Full Text PDFVet Surg
May 2017
Department of Statistics, College of Agriculture and Life Sciences, North Carolina State University, Raleigh, North Carolina.
Objective: To determine any loss of function of a single-use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries.
Study Design: Ex vivo; randomized block design (dog).
Sample Population: Carotid artery segments (n = 90) from 15 canine cadavers.
Vet Rec
April 2017
Small Animal Department, Faculty of Veterinary Medicine, Ghent University, Salisburylaan 133, Merelbeke 9820, Belgium.
The objectives of this study were to compare (1) the extent of thermal damage and (2) the time between the 5-mm LigaSure V (LS5) and 10-mm LigaSure Atlas (LS10) vessel sealing devices (VSD) when performing open ovariectomy in dogs. A prospective, randomised, clinical trial was performed in 40 client-owned sexually entire female dogs. In each dog, one ovary was randomly assigned to be surgically removed using LS5 and the contralateral using LS10.