Objectives: Focused assessment with sonography in trauma (FAST) can define life-threatening injuries in austere settings with remote real-time review by experienced physicians. This study evaluates vest-mounted microwave, satellite, and LifeLink communications technology for image clarity and diagnostic accuracy during remote transmission of FAST examinations.

Methods: Using a SonoSite, FAST was obtained on three patients with pericardial and intraperitoneal effusions and two control subjects in a remotely located U.S. Army Combat Support Hospital. A miniature vest-mounted video transmitter attached to the SonoSite sent wireless ultrasound video 20 m to a receiving antenna. The signal was then transferred over VSAT satellite systems at 512 kilobaud per second (kbps), INMARSAT satellite systems at 64 kbps, and over LifeLink on a moving ambulance through a metropolitan wireless traffic-management network. Clarity and absence or presence of effusions were recorded by 15 staff emergency physicians.

Results: Average sensitivity, specificity, and accuracy were 87% (95% confidence interval [CI]=79% to 95%), 85% (95% CI=81% to 89%), and 86% (95% CI=82% to 90%) for the Premier Wireless Vest; 98% (95% CI=97% to 99%), 83% (95% CI=75% to 91%), and 86% (95% CI=82% to 90%) for VSAT; 95% (95% CI=94% to 96%), 70% (95% CI=58% to 82%), and 75% (95% CI=70% to 80%) for INMARSAT; and 82% (95% CI=73% to 91%), 83% (95% CI=74% to 92%), and 82% (95% CI=78% to 86%) for LifeLink with clarity of 3.0 (95% CI=2.7 to 3.3), 2.9 (95% CI=2.6 to 3.2), 1.3 (95% CI=1.2 to 1.4), and 2.1 (95% CI=1.8 to 2.4), respectively.

Conclusions: Accuracy correlated with clarity. Roaming vest transmission of FAST provides interpretable, diagnostic imagery at the distances used in this study. VSAT provided the best clarity and diagnostic value with the lighter, more portable INMARSAT serving a lesser role for remote clinical interpretation. LifeLink performed well, and further infrastructure improvements may increase clarity and accuracy.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1553-2712.2003.tb00021.xDOI Listing

Publication Analysis

Top Keywords

95%
18
focused assessment
8
assessment sonography
8
sonography trauma
8
clarity diagnostic
8
transmission fast
8
satellite systems
8
86% 95%
8
95% ci=82%
8
ci=82% 90%
8

Similar Publications

Article Synopsis
  • The study measured fibrinogen fluorescence at temperatures between 20 and 80 degrees Celsius across different pH levels.
  • It was found that raising the temperature from 20 to 40 degrees Celsius did not change the structure of fibrinogen in solutions with pH between 4.5 and 9.3.
  • However, temperatures between 40 to 50 degrees Celsius caused some structural changes in neutral solutions, and temperatures above 50-55 degrees Celsius led to significant denaturation of the fibrinogen molecule.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!